8  Handler Interface 
The instrument is equipped with a 9 pin Handler interface for external control and outputting 
measurement and sorting results. 
Terminals and Descriptions: 
Mating connector: Amphenol PN: L77S-DE09S Hood: 17E-1724-1 
Pin 
IO 
Name 
Description 
1 
In 
TRIG 
Trigger signal, falling edge is valid. When the 
instrument is in the EXTernal trigger mode and 
this signal is valid, the instrument will trigger a 
measurement. 
2 
Out 
PASS2 (P2) 
Low Bin 2 comparison result output 
3 
Out 
FAIL 
The low bin comparison result output signal will 
be valid. 
4 
Power 
+5V 
Internal 5V power output 
5 
Power 
EXT_VCC 
When Handler VCC is set to EXTernal, the 
output requires an external power source, +5V 
to +30V. When Handler VCC is set to INTernal, 
the pin is shorted to the internal 5V power 
output. 
6 
Out 
PASS1 (P1) 
Low Bin 1 comparison result output 
7 
Out 
PASS3 (P3) 
Low Bin 3 comparison result output 
8 
Out 
EOC 
Measurement end signal. 
9 
Power 
EXT_GND 
When the “Handler” power supply is set as 
“External”, the terminal is external GND; when 
the “Handler” power supply is set as “Internal”, 
the terminal is internal GND. 
Table 7 - Handler Pinout 