User Manual
42 
3.  Use the keypad to enter the delay value. Manual trigger delay range is: 0 ms to 9999 ms. 
4.1.5  AVERAGE 
Press AVERAGE to open a numeric keypad. The user can set the average number a samples 
taken per trigger, ranging from 1 to 255. The larger the sample sizes reduces variation at the 
cost of a longer time to display the result. 
4.1.6  SPEED 
This function is identical to SPEED on <MEAS DISP> page. See 3.1.4 for details. 
4.1.7  0 ADJ 
Select the zero adjustment offset state. OFF will not include the offset in the measurement 
calculation. ON will include the offset in the measurement calculation. The offset value used for 
this setting is calculated after the cable calibration is completed. See 2.5 for details. 
  0 ADJ OFF: measured value + cable resistance will be displayed. 
  0 ADJ ON: measured value - cable resistance will be displayed. 
Note: The Cable Calibration must be competed correctly before this future can be used 
4.2  Bin Setup 
The BIN SETUP page is where the user can define the parameters for the 3 bins. 
To access the page press the SETUP button and select BIN SETUP in menu options. 
Figure 26 - Bin Setup 
Note: If delay time is set to 0 ms, the contact settling time detection cannot be executed. It is recommended 
to set the delay time to be greater than 1 ms. 