NOTE: Variables in the above table:  
n
: The total number of samples. 
x
: Measurement results of each sample measurement. The data are saved in the instrument buffer memory.  
USL
: Upper specification limit. If NOM, % is 100±10% then 
USL
= 110. 
LSL
: Lower specification limit. If NOM, % is 100±10% then 
LSL
 = 90. 
When 
Cp, Cpk > 1.33
, the working capacity is ideal. 
When 
1.33  Cp, Cpk >1.00
, the working capacity is sufficient. 
When 
1.00  Cp, Cpk
, the working capacity is insufficient. 
  Hi: Incremented when the measurement result exceeds the upper limit value 
USL
. 
  Lo: Incremented when the measurement result is less than the lower limit value 
LSL
. 
  In: Incremented when the measurement result is within the limits. 
  Max: Maximum measurement result among the current data set. 
  MaxIndex: Sample index number that corresponds to the maximum measurement 
result. 
  Min: Minimum measurement result among the current data set.  
  MinIndex: Sample index number that corresponds to the minimum measurement 
result. 
  R: Measured resistance. 
  num: Incremented every trigger, even when the source terminal is open. 
  v: Incremented for every measurement recorded 
Example: NOM, % is 100±10% 
Sample Index 
1 
2 
3 
4 
5 
6 
8 
Result (Ω) 
100.4 
101.6 
103.7 
98.4 
87.9 
112.1 
86.5 
Results: 
Hi = 1, Lo = 2, In = 5, Max = 112.1, MaxIndex = 6, Min = 86.5, MinIndex = 8 