User Manual

39
NOTE: Variables in the above table:
n
: The total number of samples.
x
: Measurement results of each sample measurement. The data are saved in the instrument buffer memory.
USL
: Upper specification limit. If NOM, % is 100±10% then
USL
= 110.
LSL
: Lower specification limit. If NOM, % is 100±10% then
LSL
= 90.
When
Cp, Cpk > 1.33
, the working capacity is ideal.
When
1.33 Cp, Cpk >1.00
, the working capacity is sufficient.
When
1.00 Cp, Cpk
, the working capacity is insufficient.
Hi: Incremented when the measurement result exceeds the upper limit value
USL
.
Lo: Incremented when the measurement result is less than the lower limit value
LSL
.
In: Incremented when the measurement result is within the limits.
Max: Maximum measurement result among the current data set.
MaxIndex: Sample index number that corresponds to the maximum measurement
result.
Min: Minimum measurement result among the current data set.
MinIndex: Sample index number that corresponds to the minimum measurement
result.
R: Measured resistance.
num: Incremented every trigger, even when the source terminal is open.
v: Incremented for every measurement recorded
Example: NOM, % is 100±10%
Sample Index
1
2
3
4
5
6
8
Result ()
100.4
101.6
103.7
98.4
87.9
112.1
86.5
Results:
Hi = 1, Lo = 2, In = 5, Max = 112.1, MaxIndex = 6, Min = 86.5, MinIndex = 8