Parameter
Description
Baud rate
9600, 19200, 28800, 38400, 48000, 57600, and 115200.
Parity and
data bit
None/8 bits
Stop bit
1
Flow control
None
The RS232C interface does not support hardware flow control (only transmit, receive,
and ground pins are used). The programmer should be aware of this communication
error. Therefore, limitation and notice the command process time of the instrument.
If the remote commands are sent too fast to the instrument, the internal buffer may
overrun and cause a adding a delay between commands may be necessary to allow
time for the meter to process.
7.2 USBTMC
USB The standard USB port is a USBTMC-compliant port that can be used for remote
communication. There are no additional settings in the menu system for USB configuration. The
only requirement is that the USBTMC driver be installed. It is included when a VISA software is
installed on the computer. (We recommend using NI-VISA, which can be downloaded at
http://www.ni.com/visa/).
7.3 USB (USBCDC - Virtual COM)
The standard USB port is a virtual COM port that can be used for remote communication. There
are no settings in the menu system for USB configuration. The settings are the same as the
settings for RS232C.
The USB interface does not support hardware flow control (only transmit, receive,
and ground pins are used). The programmer should be aware of this limitation and
notice the command process time of the instrument. If the remote commands are
sent too fast to the instrument, the internal buffer may overrun and cause a
communication error. Therefore, adding a delay between commands may be
necessary to allow time for the meter to process.