User Manual
126 
Parameter 
Description 
Baud rate 
9600, 19200, 28800, 38400, 48000, 57600, and 115200. 
Parity and 
data bit 
None/8 bits 
Stop bit 
1 
Flow control 
None 
The RS232C interface does not support hardware flow control (only transmit, receive, 
and ground pins are used). The programmer should be aware of this communication 
error. Therefore, limitation and notice the command process time of the instrument. 
If the remote commands are sent too fast to the instrument, the internal buffer may 
overrun and cause a adding a delay between commands may be necessary to allow 
time for the meter to process. 
7.2  USBTMC 
USB The standard USB port is a USBTMC-compliant port that can be used for remote 
communication. There are no additional settings in the menu system for USB configuration. The 
only requirement is that the USBTMC driver be installed. It is included when a VISA software is 
installed on the computer. (We recommend using NI-VISA, which can be downloaded at 
http://www.ni.com/visa/).  
7.3  USB (USBCDC - Virtual COM) 
The standard USB port is a virtual COM port that can be used for remote communication. There 
are no settings in the menu system for USB configuration. The settings are the same as the 
settings for RS232C. 
The USB interface does not support hardware flow control (only transmit, receive, 
and ground pins are used). The programmer should be aware of this limitation and 
notice the command process time of the instrument. If the remote commands are 
sent too fast to the instrument, the internal buffer may overrun and cause a 
communication error. Therefore, adding a delay between commands may be 
necessary to allow time for the meter to process. 










