User Manual

114
Figure 58 External Trigger Timing
Figure 59 Internal Trigger Timing
4.4.8 Err.OUT signal
The Err.OUT signal can be found on pin 11 of the [Handler]. A measurement error occurs when
the instrument has lost contact with a DUT. Press Err.OUT to select between:
SYNC: If the measurement error is detected during t2, the instrument will output an
error signal. No error detection will be taken in other time ranges.
ASYNC: If the measurement error is detected during t2, the instrument will output a
measurement error signal. If the measurement error is detected during t2 and t3 (time
until next trigger) and lasts for over 5 ms, the instrument will output an error signal. If
the measurement error disappears within 5 ms, no measurement error signal will be
output.