User Manual
114 
Figure 58 External Trigger Timing 
Figure 59 Internal Trigger Timing 
4.4.8  Err.OUT signal 
The Err.OUT signal can be found on pin 11 of the [Handler]. A measurement error occurs when 
the instrument has lost contact with a DUT. Press Err.OUT to select between: 
  SYNC: If the measurement error is detected during t2, the instrument will output an 
error signal. No error detection will be taken in other time ranges. 
  ASYNC: If the measurement error is detected during t2, the instrument will output a 
measurement error signal. If the measurement error is detected during t2 and t3 (time 
until next trigger) and lasts for over 5 ms, the instrument will output an error signal. If 
the measurement error disappears within 5 ms, no measurement error signal will be 
output. 