User Manual
100 
  OFF: When OFF is selected, the statistic measurements are disabled. All other functions 
and buttons are enabled. 
3.4.4  Statistical Analysis Parameters 
Parameter 
Variable 
Description 
Formula 
MEAN 
 
Average value 
STDEV 
 
Population 
Standard 
Deviation 
 
 
SaSTDEV 
Sample 
Standard 
Deviation 
 
 
 
Cp 
Process 
Capability Index 
(Dispersion) 
 
CpK 
Process 
Capability Index 
(Deviation) 
 
   
Table 10 Statistical Analysis Parameters 
NOTE: Variables in Table 2 - Statistical Analysis Parameters formulas: 
n
: The total number of samples. 
x
: Measurement results of each sample measurement. The data are saved in the instrument buffer memory.  
USL
: Upper specification limit. If NOM, % is 100±10% then 
USL
= 110. 
LSL
: Lower specification limit. If NOM, % is 100±10% then 
LSL
 = 90. 
When 
Cp, Cpk > 1.33
, the working capacity is ideal. 
When 
1.33  Cp, Cpk >1.00
, the working capacity is sufficient. 
When 
1.00  Cp, Cpk
, the working capacity is insufficient. 
  Hi: Incremented when the measurement result exceeds the upper limit value 
USL
.  
  Lo: Incremented when the measurement result is less than the lower limit value 
LSL
. 
  In: Incremented when the measurement result is within the limits. 
  Max: Maximum measurement result among the current data set. 
  MaxIndex:  Sample  index  number  that  corresponds  to  the  maximum  measurement 
result. 