User Manual

100
OFF: When OFF is selected, the statistic measurements are disabled. All other functions
and buttons are enabled.
3.4.4 Statistical Analysis Parameters
Parameter
Variable
Description
Formula
MEAN
Average value
STDEV
Population
Standard
Deviation
󰇛 󰇜

SaSTDEV

Sample
Standard
Deviation

󰇛 󰇜

Cp
Process
Capability Index
(Dispersion)
 


CpK

Process
Capability Index
(Deviation)

 
  


Table 10 Statistical Analysis Parameters
NOTE: Variables in Table 2 - Statistical Analysis Parameters formulas:
n
: The total number of samples.
x
: Measurement results of each sample measurement. The data are saved in the instrument buffer memory.
USL
: Upper specification limit. If NOM, % is 100±10% then
USL
= 110.
LSL
: Lower specification limit. If NOM, % is 100±10% then
LSL
= 90.
When
Cp, Cpk > 1.33
, the working capacity is ideal.
When
1.33 Cp, Cpk >1.00
, the working capacity is sufficient.
When
1.00 Cp, Cpk
, the working capacity is insufficient.
Hi: Incremented when the measurement result exceeds the upper limit value
USL
.
Lo: Incremented when the measurement result is less than the lower limit value
LSL
.
In: Incremented when the measurement result is within the limits.
Max: Maximum measurement result among the current data set.
MaxIndex: Sample index number that corresponds to the maximum measurement
result.