User Manual

Chapter 7
TEMPerature
The following commands are available on the 2841 only.
7.1 TEMPerature:CORRect:STATe
Description The state of the temperature correction function. Sets the TC/t mode
to TC.
Note Turning the temperature correction function on stops the temperature
conversion function.
Command Syntax TEMPerature:CORRect:STATe <bool>
Parameters on, 1, o, 0
Example temp:correct:state on
ery Syntax TEMPerature:CORRect:STATe?
Returned Parameters 1 or 0 (on and o respectively)
7.2 TEMPerature:CORRect:PARameter
Description Sets the “reference temperature” and “temperature coeicient” of the tem-
perature correction function.
Command Syntax TEMPerature:CORRect:PARameter <nr2>,<nr1>
Parameters <Reference Temp.>,<Temp. Coeicient>
The reference temperature range is -10.0 to 99.9
C
The temperature coeicient range is -99999 to 99999 ppm/
C
Example temp:corr:par 25,3390
ery Syntax TEMPerature:CORRect:PARameter?
Returned Parameters <nr2>,<nr1>
14