To change the time base for the Zoom window, turn the Horizontal Scale knob. The Horizontal 
Scale knob controls the size of the darker zoom window on the upper waveform. The Horizontal 
Position knob sets the position of the zoom window. Negative delay values indicate that a portion 
of the waveform before the trigger event is displayed, and positive values indicate a portion of 
the waveform after the trigger event. 
If the zoomed time/div (prefaced with Z at the top of the screen) is substantially smaller than the 
main time/div, you may see significant jitter on things like the edges of fast-rising pulses. You can 
get  a  stable  display  for  making  measurements  by  pressing  the  Single  trigger  button. Turn 
persistence on to measure jitter. 
To change the zoomed time base of the normal window, turn off the Zoom function and adjust 
the Horizontal Scale knob. 
3.5  Run Control 
Figure 21 - Run control buttons 
The Auto Setup button automatically adjusts the oscilloscope's settings to get 
a stable display. 
Use this button to set the state of the instrument to RUN or STOP. In the RUN 
state, the button glows yellow; in STOP state, the button glows red. 
Press the button to reset the oscilloscope to its default setup state. This is a 
useful starting point to let you manually adjust the oscilloscope for your 
needs. 
This button has two effects. When measurement statistics are being displayed, 
pressing this button will set the statistics to zero and start accumulating data 
again. When screen persistence is turned on, pressing this button will clear the 
persisted waveforms. 