User Manual
138 
b.  Turn the Universal Knob to select the channel connected to the SPI serial clock line. 
c.  Press  the  Threshold  softkey  and  turn  the  Universal  Knob  to  select  the  clock  signal 
threshold voltage level. The threshold voltage level is used in decoding and it will become 
the trigger level when the trigger type is set to the selected serial decode slot. 
d.  Press the Edge Select softkey to select rising edge or falling edge for the selected clock 
source. This determines which clock edge the oscilloscope will use to latch the serial data. 
6.  Press the   softkey to return to the SIGNAL menu. 
7.  Press the MISO softkey to enter the SPI MISO menu. 
In the MISO menu: 
a.  Press  the  MISO  softkey  and  turn  the  Universal  Knob  to  select  the  channel  that  is 
connected to a second SPI  serial data line. If the channel selected is off, it should be 
switched on. You can also select Disable. 
b.  Press  the  Threshold  softkey  and  turn  the  Universal  Knob  to  select  the  MISO  signal 
threshold voltage level. The threshold voltage level is used in decoding and it will become 
the trigger level when the trigger type is set to the selected serial decode slot. 
8.  Press the   softkey to return to the SIGNAL menu. 
9.  Press the MOSI softkey to enter the SPI MOSI menu. 
Figure 112 - MOSI menu 
In the MOSI menu: 
a.  Press  the  MOSI  softkey  and  turn  the  Universal  Knob  to  select  the  channel  that  is 
connected to a SPI serial data line. If the channel selected is off, it should be switched on. 
You can also select Disable. 
b.  Press  the  Threshold  softkey  and  turn  the  Universal  Knob  to  select  the  MOSI  signal 
threshold voltage level. The threshold voltage level is used in decoding and it will become 
the trigger level when the trigger type is set to the selected serial decode slot. 
10. Press the   softkey to return to the SIGNAL menu. 
11. Press the CS softkey to open the SPI CS menu. 
Figure 113 - SPI CS Menu 
In the CS menu: 
a.  Press the  Cs  Type  softkey  to  select a  framing signal  that the  oscilloscope  will use to 
determine which clock edge is the first clock edge in the serial stream. You can set the 
oscilloscope to trigger during a high chip select (CS), a low chip select (~CS), or after a 
Timeout period during which the clock signal has been idle. 