User Manual
20 
User Interface Functionality 
1. Operating state 
The states are Arm, Ready, Trig’d (triggered), Stop, Auto. 
2. Horizontal Timebase 
Represents the time per division on the horizontal axis. Turning the horizontal scale knob (the 
left knob in the Horizontal control area) changes the time per division setting from 1 ns/div to 50 
s/div. 
3. Trigger position parameter (delay) 
Shows the time difference between a trigger point and the center of the screen. Turn clockwise 
or  counterclockwise  to  make  the  waveform  move  right  or  left,  which  will  cause  the  delay 
parameter to decrease or increase, respectively. Press the horizontal position knob to reset the 
delay parameter to zero (the trigger position mark will then be in the middle of the screen). 
4. Trigger position mark 
Displays the trigger point on the waveform. The delay parameter is zero at this point. 
5. Frequency counter  
Displays the frequency of the trigger source waveform. 
6. Sample rate/memory depth  
Displays the current sample rate (Sa) and memory depth (Curr) of the oscilloscope. Use the 
horizontal scale knob to modify the parameters. 
7. Trigger settings 
The trigger settings are always displayed on the upper-right side of the screen. 
Figure 5 - Trigger settings display 