User Manual
95 
Window 
Characteristics 
Applications 
Rectangle 
The best frequency 
resolution, but the poorest 
amplitude resolution. Similar 
to when no window is 
applied. 
Transient or short pulses. Sine 
waveform with the same amplitude 
and rather similar frequencies. Wide 
band random noise with relatively 
slowly changing waveform spectrum. 
Hanning 
Better frequency resolution; 
poorer amplitude resolution. 
Sine, periodic and narrow band 
random noise. 
Hamming 
Somewhat better frequency 
resolution than Hanning. 
Transient or short pulse, the signal 
levels before and after the 
multiplication are rather different. 
Blackman 
The best amplitude 
resolution; the poorest 
frequency resolution. 
Single frequency signal, search for 
higher order harmonics. 
Table 7 - FFT window instructions 
5.  Press the Center softkey and turn the  Universal Knob to set the frequency of  the  center 
vertical line. 
6.  Press the Hz/div softkey and turn the Universal Knob to set the vertical range of the display. 
7.  Press the FFT Zoom softkey and turn the Universal Knob or turn the horizontal scale knob to 
select the desired magnification (1X, 2X, 5X, 10X). Set the FFT Zoom to a higher magnification 
to see more details of the FFT waveform. 
8.  Press the Scale softkey to select the units per division for the vertical axis. The units of the 
vertical axis  can be dBVrms  (logarithmic) or Vrms  (linear). If you need  to display the  FFT 
frequency spectrum with a large dynamic range, dBVrms is recommended. 
9.  Press the Reference Level softkey and turn the Universal Knob to adjust the position of the 
FFT's vertical position. 
10. Press the Unit softkey and turn the Universal Knob to display with a Vrms (linear) or dBVrms 
(logarithmic) vertical scale. 
11. Press the Display softkey to select Split, Full Screen, and Exclusive display mode. 
  Split: the source channel and the FFT operation results are displayed separately. The time 
domain and frequency domain signals are displayed on separate portions of the screen. 
  Full Screen: the source channel and the FFT operation results are displayed in the same 
window  to  view  the  frequency  spectrum  more  clearly  and  to  perform  more  precise 
measurements. 
  Exclusive: only the FFT operation results are displayed in full screen. 