Pulse Trigger 
This trigger is used to trigger on a positive or negative pulse with a specified width. 
To set the Pulse Trigger: 
1.  Press the Setup button to enter the TRIGGER menu. 
2.  Press the Type softkey. Turn the Universal Knob to select Pulse and push the knob to confirm. 
3.  Press the Source softkey. Turn the Universal Knob to choose which analog channel to use for 
the trigger source. 
4.  Turn the Trigger Level Knob to adjust the trigger level. The pulse width will be measured at 
this voltage on the waveform. 
5.  Press  the  Polarity  softkey  to  trigger  on  a  Positive  or  Negative  pulse.  The  current  trigger 
polarity is displayed at the upper right corner of the screen. 
6.  Press the Limit Range softkey. Turn the Universal Knob to select the desired condition. 
  < (Less  than  specified  time):  Trigger when the  measured pulse width  is less than the 
specified time. 
For example, for a positive pulse, if you set the pulse width < 100 ns, the oscilloscope will trigger 
on the following waveform: 
Note:  The  lower  trigger  level  cannot  be  larger  than  the  upper  trigger  level.  In  the  trigger  state 
message box, L1 means the upper trigger lever while L2 means the lower trigger level. 
Positive Pulse Width 
Negative Pulse Width 
Trigger Level 
Figure 52 - Pulse Trigger 