User Manual
29 
Figure 13 – Oscilloscope probe 
Connect the probe to the oscilloscope and connect the ground terminal to ground before you 
take any measurements. 
WARNING: SHOCK HAZARD 
To avoid electric shock when using the probe, please make certain that the insulated wire of 
the probe is in good condition and do not touch the metallic parts of the probe when it is 
connected to a high voltage. 
Do not connect the probe's ground lead (e.g., using the supplied alligator clip) to any point in 
a  circuit that  is  not  at  ground  potential.    If  you're  not  sure  whether a  point  you  want  to 
connect to is at ground potential, first check it with a known-working high-impedance digital 
voltmeter. 
Probe Attenuation 
Probes have attenuation factors that can affect the vertical display of the signal. Before making 
measurements with the probe, verify the probe's attenuation matches the attenuation setting of 
the oscilloscope channel it is connected to. Push the numbered button corresponding to the 
channel the probe is connected to and ensure the attenuation factor shown on the Probe softkey 
matches that of the probe. Failure to do this will result in significant measurement error. 