User Manual
137 
  Red dots in the decode line indicate that more data can be displayed. Scroll or expand the 
horizontal scale to view the data. 
Interpreting I2C Lister Data 
In addition to the standard Time column, the I2C Lister contains these columns: 
  NO: frame number from left to right. 
  TIME: time for each frame. 
  ADDRESS: colored blue for writes, yellow for reads. 
  R/W: yellow R for read, dark green W for write and black X for missing. 
  DATA: data bytes. 
15.2  SPI Serial Decode 
The following  order  of operations should be followed to ensure that the  trigger and decode 
functions are properly set: 
 Setup for SPI Signals 
 SPI Decode 
Setup for SPI Signals 
Serial Peripheral Interface (SPI) signals setup consists of connecting the oscilloscope to a clock, 
MOSI data, MISO data, and framing signal, then setting the threshold voltage level for each input 
channel, and finally specifying any other signal parameters. 
To set up the oscilloscope to capture SPI signals: 
1.  Press the Decode button to enter the DECODE menu. 
2.  Press the Decode softkey to select the desired slot (Decode 1 or Decode 2). 
3.  Press  the  Protocol softkey and select  SPI with the  Universal Knob  and press the knob to 
confirm. 
4.  Press the Signal softkey to enter the SPI SIGNAL menu. 
Figure 110 - SPI SIGNAL menu 
5.  Press the CLK softkey to enter CLK (clock) menu. 
Figure 111 - SPI CLK menu 
In the SPI CLK menu: 
a.  Press the CLK softkey. 