  Start - The oscilloscope triggers when SDA transitions from high to low while the SCL 
clock is high. For triggering purposes (including frame triggers), a restart is treated as a 
start condition. 
  Stop - The oscilloscope triggers when data (SDA) transitions from low to high while the 
clock (SCL) is high. 
Figure 69 – Start vs stop condition 
  Restart - The oscilloscope triggers when another start condition occurs before a Stop 
Condition. 
  No Ack -The oscilloscope triggers when SDA is high during any Ack SCL clock bit. 
  EEPROM -The trigger looks for the EEPROM control byte value 1010xxx on the SDA line, 
followed by a Read bit and an Ack bit. It then looks for the data value and qualifier set by 
the Limit Range softkey and the Data1 softkey. When this event occurs, the oscilloscope 
will trigger on the clock edge for the Ack bit after the data byte. This data byte does not 
need to occur directly after the control byte. 
Figure 70 - EEPROM trigger 
To set the oscilloscope to trigger on an EEPROM Data Read condition: 
a)  Press the Limit  Range  softkey to set the oscilloscope to trigger when  the data is 
= (equal to), < (less than), or > (greater than) the data value set in the Data1 softkey. 