User Manual
36 
Specify Probe Attenuation Factor 
Set the channel's attenuation factor to match the attenuation  of the probe you are using to 
ensure correct voltage or current measurements. To do this, press the 1 button on the front 
panel,  then  press  the  Probe  softkey  and  turn  the  Universal  Knob  to  select  the  desired 
attenuation. Push the knob to select the chosen value. The default is 1X. You can also press the 
Probe softkey repeatedly to change the channel's probe attenuation factor. 
Specify Channel Input Impedance 
Select the channel input impedance: 1 MΩ and 50 Ω. Do not exceed a voltage input of more than 
5 V absolute value for a 50 Ω impedance input or 400 V absolute value for a 1 MΩ impedance 
input. A high impedance of 1 MΩ minimizes loading the device under test. 
Press the channel button on the front panel, then press the Impedance softkey to toggle between 
1 MΩ and 50 Ω. The default is 1 MΩ. The channel's input impedance is displayed in the channel 
label at the right side of the screen. 
Specify the Amplitude Unit 
You can display the channel's measurement unit as volts (V) or amperes (A). When the unit is 
changed, the unit displayed in the channel label will change accordingly. The default setting is V. 
1.  Press the 1 button on the front panel to display the CH1 menu. 
2.  Press the Next Page softkey. 
3.  Press the Unit softkey to select V or A. 
Deskew 
Use the Deskew softkey to time-coordinate probe measurements, as they can have small delays 
that can result in significant power waveform errors.  The adjustment ranges from -100 to 100 
ns.  A common use is to reduce cable length induced delay. Make these adjustments when you 
first  connect  the  probes,  then  subsequently  when  the  measurement  hardware  or  the 
temperature change. 
Inverting a Waveform 
When Invert is set to On, the voltage of each measured point is multiplied by -1, which inverts 
the  waveform.  Note  this  also  multiplies  the  trigger  voltage  by  -1  so  that  a  stable  display  is 
maintained. Inverting a channel also affects the results of math functions and measure functions. 
1.  Press the 1 button on the front panel to display the CH1 menu. 
2.  Press the Next Page softkey to enter the second page of the CH1 function menu. 
3.  Press the Invert softkey to turn on or off inverting display. 