17.4  Set Default Values 
A softkey in the WAVEFORM menu lets you restore the default wave type, frequency, amplitude, 
and DC offset: 
1.  Press the Wave Gen button to enter the WAVEFORM menu. 
2.  Press the Setting softkey to enter the SETTING menu. 
3.  Press the Default softkey to restore the default values a sine waveform of 1 kHz frequency, 4 
Vpp, and 0 V DC offset. 
17.5  AWG Self Cal 
If the ambient operating temperature changes more than 5 °C and the instrument runs more 
than  30  minutes,  you  should  run  the  AWG  self  calibration  routine  to  reduce  thermal  offset 
effects. 
1.  Press the Wave Gen button to enter the WAVEFORM menu. 
2.  Press the Setting softkey to enter the SETTING menu. 
3.  Press the AWG Self Cal softkey to run the AWG self-calibration program. 
Figure 144 - AWG self-calibration 
4.  When the self calibration program completes (it will take approximately 5 seconds), press the 
Run/Stop button to return to the SETTING menu. 