User Manual
123 
Figure 99 - Virtual panel 
Auxiliary Output 
You can set the type of the signal output from the TRIGGER OUT (also labelled PASS/FAIL) BNC 
connector on the rear panel. 
1.  Press the Utility button to enter the UTILITY menu. 
2.  Press the I/O softkey to enter the I/O SET function menu. 
3.  Press the Aux Output softkey to select Trig Out or Pass/Fail. The default setup is Trig Out. 
  Trig  Out:  The  oscilloscope  outputs  a  positive-going  pulse  at  each  trigger  event.  The 
maximum signal frequency will be 140 kHz. 
  Pass/Fail:  The  oscilloscope  will  output  a  pulse  signal  when  failed  waveforms  are 
detected. 
The output pulse on the TRIGGER OUT BNC connector on the back will be nominally a 3.3 positive 
pulse about 600 ns wide for each trigger event. 
11.7  Quick-Cal 
Quick-Cal automatically calibrates the  oscilloscope when  ambient temperature variations are 
affecting  the  measurement  accuracy.  If  the  environmental  temperature  of  the  oscilloscope 
working area is not stable, follow these steps: 
1.  Press the Utility button to enter the UTILITY menu. 
2.  Press the Next Page softkey to go to the second page of the UTILITY function menu. 
3.  Press the Quick-Cal softkey until it reads On. This enables Quick-Cal. 