User Manual

50
Trigger on Rising edge and Falling edge of the
trigger signal.
Mode
Auto
Use this mode to let the acquisition free-run in
the absence of a valid trigger; This mode allows
an untriggered, scanning waveform at 100
ms/div or slower time base settings.
Normal
Use this mode when you want to see only valid
triggered waveforms; when you use this mode,
the oscilloscope does not display a waveform
until after the first trigger.
Single
When you want the oscilloscope to acquire a
single waveform, press the “SINGLE” button.
Set up
Enter the “Trigger Setup Menu” (See Table 3.17).
Table 3.17 Trigger Setup Function Menu
Option
Setting
Explain
Coupling
DC
Passes all components of the signal
AC
Blocks DC components, attenuates
signals below 50 Hz.
HF Reject
Attenuates the high-frequency
components above 150 kHz.
LF Reject
Blocks the DC component, attenuates the
low-frequency components below 7 kHz.
Holdoff
Using the “universal” knob to adjust
holdoff time (sec), the holdoff value is
displayed.
Holdoff
Reset holdoff time to 100ns.