Safety Summary The following safety precautions apply to both operating and maintenance personnel and must be followed during all phases of operation, service, and repair of this instrument. Before applying power to this instrument: • Read and understand the safety and operational information in this manual. • Apply all the listed safety precautions. • Verify that the voltage selector at the line power cord input is set to the correct line voltage.
Electrical Power This instrument is intended to be powered from a CATEGORY II mains power environment. The mains power should be 115 V RMS or 230 V RMS. Use only the power cord supplied with the instrument and ensure it is appropriate for your country of use. Ground the Instrument To minimize shock hazard, the instrument chassis and cabinet must be connected to an electrical safety ground. This instrument is grounded through the ground conductor of the supplied, three-conductor AC line power cable.
Do not operate instrument if damaged If the instrument is damaged, appears to be damaged, or if any liquid, chemical, or other material gets on or inside the instrument, remove the instrument’s power cord, remove the instrument from service, label it as not to be operated, and return the instrument to B&K Precision for repair. Notify B&K Precision of the nature of any contamination of the instrument.
Do not substitute parts that are not approved by B&K Precision or modify this instrument. Return the instrument to B&K Precision for service and repair to ensure that safety and performance features are maintained. For continued safe use of the instrument • Do not place heavy objects on the instrument. • Do not obstruct cooling air flow to the instrument. • Do not place a hot soldering iron on the instrument. • Do not pull the instrument with the power cord, connected probe, or connected test lead.
Contents 1 IMPORTANT SAFETY INSTRUCTIONS 1.1 Precautions 7 7 2 Product Overview 2.1 Package Contents 2.2 Power Requirements 2.3 Interface and Controls 8 8 8 8 3 Basic Operation 3.1 Setting Voltage and Current by Rotary Knob and Up/Down Keys 3.2 Setting Voltage and Current Using Keypad 9 9 9 4 Operating Instructions 4.1 Output Enable 4.2 Output Disable 4.3 Locking the Keypad and Rotary Knob 4.4 Unlocking the Keypad and Rotary Knob 4.5 Setting the RS-485 Address 4.6 Setting the Upper Voltage Limit 4.
IMPORTANT SAFETY INSTRUCTIONS 1. Do not use this apparatus near water. 2. Clean only with dry cloth. 3. Do not block any ventilation openings. 4. Do not install unit near any heat source or heating emitting devices. 5. Prevent the power cord from being walked on or pinched. 6. Unplug this unit during lightning storms or when unused for long periods of time. 1.1 Precautions 1. The unit must be used within its specified range. The rated input voltages can be found on the rating label at the back of the unit.
Product Overview 2.1 Package Contents • Power Supply (1696B, 1697B or 1698B) • Power cord • USB cable 2.2 Power Requirements Parameter Value Voltage 100 - 240 VAC Frequency 50/60 Hz Fuse 4A Slow Blow 250V 2.
Basic Operation 3.1 Setting Voltage and Current by Rotary Knob and Up/Down Keys 1. Press to switch between setting voltage and current. 2. Rotate control knob toggle the cursor position. or press and to set voltage/current level. Press the control knob to 3.2 Setting Voltage and Current Using Keypad 1. Press to switch between setting voltage and current. 2. Use the numeric keypad to to input voltage or current value. 3. Press Enter to confirm input values.
Operating Instructions 4.1 Output Enable 1. Press . 2. Press within 3 seconds to ENABLE output . within 3 seconds to DISABLE output . 4.2 Output Disable 1. Press . 2. Press 4.3 Locking the Keypad and Rotary Knob 1. Press 2. Press . within 3 seconds to lock keypad and rotary knob . 4.4 Unlocking the Keypad and Rotary Knob 1. Press 2. Press within 3 seconds to unlock keypad and rotary knob . 4.5 Setting the RS-485 Address 1. Press 2. Press 3. Use numeric keypad 4.
1. Press . 2. Press on the numeric keypad within 3 seconds to enter upper voltage limit adjustment menu. The first line says “Over” and the second if the voltage limit setting. . 3. Use the numeric keypad to 4. Press to input upper voltage limit. to confirm. 4.7 Enable Output at Power Up This feature limits the upper level setting of output voltage to prevent inadvertent setting of high voltage, which may damage your application.
1. Press 2. Press . on numeric keypad within 3 seconds to enter into SCPI enable/disable menu. 3. Use the rotary knob to toggle between ‘Y’ to enable SCPI and ‘N’ to disable SCPI command protocol and use the extended protocol command set instead. 4. Press to confirm.
Using Programming Features 5.1 Preset Program 1. Press Program . 2. Use the numeric keypad 3. Use 4. Press thru and to select a preset voltage and current value. to adjust preset voltage and current values. within 3 seconds to confirm or press to exit program setting. 5.2 Setting Timed Program The unit can be programmed to operate up to 20 timed subprograms (0-19 STEP as shown in the display).
Maintenance 6.1 Recalibration The purpose of recalibration is to reduce the difference between the set values and the displayed values on the LCD display. Recalibration is only required when this difference is greater than 0.1 V for voltage or -0.01 A / +0.02 A for current. 6.2 Troubleshooting Keypad and Dial do not work Check key lock symbol for lock state, unlock unit by pressing then lem persists.
PC Software Control and Installation The PC software provides remote communication, data logging, front panel emulation, and timed programming capabilities using the USB or RS-485 interfaces. The software supports Windows 7, 8, 8.1, 10. Note: Do not connect both USB and RS-485 simultaneously. 7.1 USB Driver Installation Remote communication using USB requires the installation of the CP210X USB virtual COM driver available for download at bkprecision.com. The USB driver is only required for Windows 7, 8, 8.
Figure 7.3 Figure 7.4 4. Click “Next” to continue. Figure 7.4 5. Select a destination location for the software. Figure 7.5 6. Select a location for the software shortcut.Figure 7.6 7. Check the box if you would like to create a desktop shortcut and click “Next” to continue. Figure 7.7 8. Click “Install” to begin installation.Figure 7.8 7.3 Multi-Unit Control 7.3.
Figure 7.5 Figure 7.6 Also, the software uses the “SCPI” protocol, this must be enabled, see Section 7.4.2. With the instrument connected, open the PC software. Figure 7.11 shows the screen to expect when the software starts. If there is already a saved connection the software will automatically connect to the instrument. 7.4.
Figure 7.7 Figure 7.8 • Select the “Setting” tab. See Figure 7.12 • Click on “Edit” to access the connection settings. See Figure 7.13 Note: When using USB: Only connection name, connection type, and COMM port are required. See Figure 7.
Figure 7.9 Connection diagram for multiple power supply control Figure 7.10 Connection diagram between USB adapter and RS-485 connectors • Type in a connection name, select USB from the drop down, and select the correct COMM port assigned by your PC. Check the windows the device manager to determine the correct COMM port. In this case Windows has assigned COM port 12. Click “Save” and the black display will light up to indicate a successful connection. See Figure 7.
Figure 7.11 Figure 7.12 7.4.4 External Timed Program External Timed Program is completely controlled by the PC. The PC counts the step time and changes the specific voltage and current levels of the power supply. Configure up to 20 user defined voltage, current, and duration steps. 1. Click on the “External Timed Program” tab. See Figure 7.20 2. Double-click on the desired cell and use the slider to set voltage and current values. Figure 7.21 3. Set step duration with the up/down buttons.
Figure 7.13 Figure 7.14 7.4.6 Data Logging Data Log allows can be used to view present or stored output data, see Figure 7.24. Data can be saved and exported as .CSV spreadsheet file or sent to a printer. 7.4.7 Data Log Sampling Time and Voltage Upper Limit (UVL) Setting Under the “Setting” tab: • Adjust the data log sampling time. 1S means the voltage, current, and power reading will be recorded every 1 second.
Figure 7.16 Figure 7.
Figure 7.19 Figure 7.20 Figure 7.
Figure 7.22 Figure 7.23 Figure 7.24 Figure 7.
Specifications Note: All specifications apply to the unit after: 1. A temperature stabilization time of 15 minutes over an ambient temperature range of 23 ∘ C ± 5 ∘ C. 2. Short correction operation performed before making measurement. Specifications are subject to change without notice.
Programmable DC Power Supplies Models 1696B, 1697B & 1698B Specifications Note: All specifications apply to the unit after a temperature stabilization time of 30 minutes over an ambient temperature range of 23 °C ± 5 °C. Model 1696B 1697B 1698B Voltage 1 to 20 V 1 to 40 V 1 to 60 V Current 0 to 10 A 0 to 5 A 0 to 3.
Service Information Warranty Service: Please go to the support and service section on our website at bkprecision.com to obtain an RMA #. Return the product in the original packaging with proof of purchase to the address below. Clearly state on the RMA the performance problem and return any leads, probes, connectors and accessories that you are using with the device. Non-Warranty Service: Please go to the support and service section on our website at bkprecision.com to obtain an RMA #.
LIMITED ONE-YEAR WARRANTY B&K Precision Corp. warrants to the original purchaser that its products and the component parts thereof, will be free from defects in workmanship and materials for a period of one years from date of purchase. B&K Precision Corp. will, without charge, repair or replace, at its option, defective product or component parts. Returned product must be accompanied by proof of the purchase date in the form of a sales receipt.