Owner's manual

Series Mode
fCsRsDc
π
2
=
Cs Rs
RESISTANCE
Range Accuracy Resolution Test Condition
2 1%5 1m 10mArms
20 10m 10mArms
200 100m 1mArms
2K 1 0.1mArms
20K 10 10µArms
200K
1%2
100
Series
Mode
1KHz
1µArms
2M 1K
20M
2%2
10K
Parallel Mode
1KHz, 0.5Vrms
Accuracy is ± ( % of readingnumber of d igits ) an d is applied from 10 to
100% of full scale range in parallel mode measurements.
In series mode measurements, the compliance voltage (voltage drop on the
device under test) should be less than 0.2 Vrms.
INDUCTANCE
Range Accuracy Resolution Test Condition
200µH 2%2 0.1µH 10mArms, 1KHz
2mH 1µH 10mArms, 1KHz
20mH 10µH 1mArms, 1KHz
200mH
1%2
100µH
Series
Mode
0.1mArms,1KHz
2H 1mH
20H 10mH
200H
Not
Specified
100mH
Parallel Mode
120Hz, 0.5Vrms
Accuracy is ± ( % of readingnumber of digits ) When D0.1.
The accuracy is applied from 10 to 100% of full scale range in parallel mode
measurements.
Ranges for reference only.
12