User Manual

Trigger
10 2 3 4 5
List count=1 List count=2
List sequence
This list is characterized by the following current/duration pairs
Current, A Duration, ms Between times
3 1000 0 and 1
0 800 1 and 2
2 500 2 and 3
0 300 3 and 4
6 500 4 and 5
This list has 5 steps, located at the transitions 1 through 5. The first duration, 1000 ms, occurs after
the triggered event. Subsequent durations are from the previous transition to the current transition.
Here are the keystrokes to set up this list:
Keys Display
Shift + Menu :CONFIG
:LIST SET
Enter :MODE SET
Enter Select :LIST mode. This will cause the LIST annunciator to turn on.
Enter :MODE SET
:EDIT LIST FILE
Enter :CURRENT LIST
Enter Select :REPEAT.
Enter LIST COUNT= N How many steps this list has. Enter the number 5.
5 Enter COUNT 1= X.XXA Set the current for the first step. Enter 3.
3 Enter COUNT 1= X.XXMS Set the duration for the first step. Enter 1000.
1000 Enter Repeat the current and duration entry for the following 4 steps.
Enter STORE LIST FILE1 The 1 is underlined, indicating you can enter a
number to specify the "file" (i.e., block of EEPROM) to store this list to.
Choose a number via the keypad or the knob. You can choose any
number between 1-8.
Enter :EDIT LIST FILE
8500 DC Load Series Version: 030614 Page 39 of 77