Sequence for testing 6 pins ISP connector: 
1.  Insert Diagnostic POD for ISP connectors into ZIF socket 
of the programmer. Diagnostic POD must be inserted as 
40 pins device. 
2.  Interconnect 6 pins connector of Diagnostic POD with an 
ISP connector of the programmer with an ISP cable, 
included in programmer delivery package. Be sure that 
pins are interconnected properly (i.e. 1-1, 2-2, ..., 6-6). 
3.  Run selftest of ISP connector in PG4UW (Programmer /  
Selftest ISP connector). 
Sequence for testing 10 pins ISP connector: 
1.  Insert Diagnostic POD for ISP connectors into ZIF socket 
of the programmer. Diagnostic POD must be inserted as 
40 pins device. 
2. Interconnect 10 pins connector of Diagnostic POD with 
an ISP connector of the programmer with an ISP cable, 
included in delivery programmer package. Be sure that 
pins are interconnected properly (i.e. 1-1, 2-2, ..., 10-10). 
3.  Run selftest of ISP connector in PG4UW (Programmer /  
Selftest ISP connector). 
We recommend run this test every 6 months. 
Programmer / Calibration test 
Command executes test of programmer's calibration values. 
Programmer / Create diagnostic report 
Command Create Diagnostic report is used for writing more 
particular diagnostic information to Log window and 
consequently copy Log window content to clipboard. The Log 
window content can be placed from clipboard to any text 
editor. Diagnostic report is useful when error occurs in control 
program or programmer and kind of the error is, that user can 
not resolve it oneself and he must contact programmer 
manufacturer. In this case when customer send message to 