Instruction Manual
10 
5.2  Duty Cycle Control 
The DUTY CYCLE KEY can be used to alter the symmetry of a square waveform. 
Symmetry variation amounts to changing the duty cycle, a ratio of “high” to “low” 
time, effectively converting the instrument into a pulse generator. 
1.
Press the DUTY CYCLE KEY and the display will blink.  
2.
Use the number keys to input the duty cycle percentage value (between 20% - 
80% can be entered).  
3.
Press the ENTER KEY to execute duty cycle adjustment. The instrument will then 
return to the frequency display. 
4.
If you want to change the duty cycle percentage value while the instrument is 
under duty cycle mode, repeat steps 1-3 above. 
To turn off duty cycle, press the DUTY CYCLE KEY two times and the duty cycle LED 
will turn off. 
Note: The duty cycle adjustment mode only functions for a square waveform and at 
frequencies below 3 MHz. 
5.3  Function Generator Display Messages 
The LED display will show 1.000 kHz after power on. If the user inputs a frequency 
value, the LED display will blink until the user presses the ENTER KEY to execute it. If 
there is an input error, the LED display will show one of the following messages: Err-
02, Err-03, Err-04, or Err-05. The display will hold this error message for about 3 
seconds and then return to the previous display. 
Error Code 
Description 
Err-02 
Input value is too large 
Err-03 
Input value is too small 
Err-04 
Over the key waiting time, did not press ENTER KEY 
Err-05 
Duty cycle frequency is higher than 3 MHz 
Table 2 - Error Codes 