9.1 Digital Storage Oscilloscope
Specifications
All specifications are typical performance values and are not
warranted. Specifications are valid after a 30 minute warm-up
time and within ± 5°C of last “Self-Cal” temperature.
Vertical system
Scope channels 2 channels plus external trigger input
Bandwidth
60 MHz: 2540B, 2540B-GEN
100 MHz: 2542B, 2542B-GEN
Calculated rise time
(=0.35/bandwidth)
< 5.83 ns: 2540B, 2540B-GEN
< 3.50 ns: 2542B, 2542B-GEN
Coupling AC, DC and GND
BW Limit 20 MHz selectable
DC Vertical Gain
Accuracy
2 mV/div, 5 mV/div: ±4 %
10 mV/div to 5 V/div: ±3 %
DC Measurement
2 mV/div to 5 mV/div:
±(4% x reading + 0.1 x V/div + 0.5 mV)
10 mV/div to 5 V/div:
±(3% x reading + 0.1 x V/div + 1.0 mV)
Position range ±8 divisions away from the center of the screen