1.7  Probe Attenuation 
Probes are available with various attenuation factors which 
affect the vertical scale of the signal. 
You can push a vertical menu button (such as the CH 1 
button), and select the Probe option that matches the 
attenuation factor of your probe. 
Note:    The default setting for the Probe option is 1X. 
Be sure that the attenuation switch on the probe matches the 
Probe option in the oscilloscope. The included probes can 
switch between 1X and 10X. 
Note:  When the attenuation switch is set to 1X, the 
probe limits the bandwidth of the oscilloscope to 
10 MHz (according to Probe spec). To use the full 
bandwidth of the oscilloscope, be sure to set the 
switch to 10X. 
1.8  Probe Compensation 
Perform this adjustment to match your probe to the input 
channel. This should be done whenever you attach a passive 