User Manual
Performance Verification Procedure 
241 
Bandwidth 
Equipment needed: Fluke calibrator (preferred) or signal generator 
1.  Connect your calibrator’s output to CH1 of the oscilloscope 
by 50ohm feed thru connector. 
2.  Set CH1 input attenuation to 5 mV/div, DC coupling, and 
horizontal scan to 500 ns/div. 
3.  Turn on output of the calibrator for a 1 MHz sine wave. 
4.  Change the output level of calibrator until waveform height 
is 6 divisions and reading is 30 mVpp. 
5.  Record these values as a reference value. 
6.  Slowly increase frequency output of calibrator up to rated 
bandwidth of the oscilloscope. 
7.  Observe waveform and reading on screen. The size of the 
waveform should always be ≥ 4.2 divisions, and reading 
should always be ≥ 21.2 mV. 
8.  Repeat the above steps for CH2. 
Oscilloscope 
Calibrator or 
Signal Generator 