229 
Acquisition system 
Max real time 
sample rate 
1 GSa/s 
Max equivalent 
sample rate 
50 GSa/s 
Max memory depth 
(Based on Sample rate) 
1 GSa/s: 16 kpts 
500 MSa/s: 8 kpts (dual channel) 
500 MSa/s: *2.4 Mpts (single channel) 
≤ 250 MSa/s: *1.2 Mpts (single and dual channel 
operation) 
Vertical resolution  8 bits 
Sample mode  Normal, Average, Peak Detect 
Autoset 
Finds and displays all active channels, sets edge 
trigger mode on channel 1, set vertical sensitivity 
on scope channels and time base to display one 
or five periods.   
Requires minimum voltage >10 mVpp, 0.5% duty 
and minimum frequency >50 Hz. 
*Maximum number of points can only be extracted via remote 
control using the USB, RS232C, or LAN interface. 