Specifications
49
1M
(9 + 1)M
= 0.1
This is why the probe is called a 10X probe as it attenuates the signal by 10 times. You'll also see them
called X10 probes. The cable's distributed capacitance and the scope's input capacitance will shunt the
signal around R
s
at higher frequencies -- this leads to the basic fact that a passive probe's impedance
at its rated bandwidth can be more than four orders of magnitude smaller than its DC
resistance. Keep this in mind when measuring signals with high frequencies as your probe will likely be
loading your circuit significantly. A common technique to see if a probe is loading a circuit is to connect
another identical probe to the scope, then connect this second probe to where the first is probing. If the
signal from the first probe changes, then circuit loading is a significant issue.
For example, a 150 MHz probe was measured to have a distributed cable capacitance of 58 pF. With a
20 pF typical scope input capacitance, this means a capacitance of about 80 pF. At 150 MHz, the
capacitive reactance of this capacitor is 13 Ω, so you can see that this effectively shorts out the 1 MΩ
input resistance of the scope. Though the capacitive reactances are low, the capacitances still behave
as a capacitive divider and attenuate the signals appropriately. To see this is approximately true, the
adjustable capacitor is on the order of 10 pF and C
c
+ C
s
is 80 pF -- almost 90 pF. If it was 90 pF, the
reactances as impedances would have a voltage divider ratio exactly the same as the resistances (the
2 terms for the reactances are factored out and we're ignoring the much larger resistances which are
effectively open circuits).
1
90
1
90
+
1
10
= 0.1
You may wonder what the adjustable capacitor C
p
is for. This capacitance is used to compensate the
probe. Compensation means to adjust the capacitance so that the probe has the correct amplitude
response for different signal frequencies. The compensation capacitor C
p
is adjusted to make the time
constant R
p
C
p
equal to the time constant of the distributed cable capacitance and scope's capacitance
of R
s
(C
c
+ C
s
). This yields a circuit with minimal distortion of the signal. A typical passive probe has an
adjustment between 10 to 30 pF.
For higher frequency passive probes, the equivalent circuit may be more complex -- and compensation
may be more complicated than simply adjusting a single capacitor. In fact, probes and their circuits
need to be analyzed as transmission lines -- and such analysis will show that probe design is not a
trivial task.
You can measure your probe's DC attenuation with a digital multimeter, but you'll have to calculate the
probe's voltage drop from knowing the digital multimeter's input resistance. For example, with a 10X
probe and a digital multimeter with a measured 10.05 MΩ input resistance, an 11.44 volt DC signal was
measured with a probe. The digital multimeter read 6.05 volts using the probe. The relevant voltage
divider is the 10 MΩ input resistance of the digital multimeter in series with the 9 MΩ of the probe.
Thus, the calculated voltage that the digital multimeter should have measured is:
10.05
9 + 10.05
11.44 = 6.035 volts
This probe's measured attenuation was within 0.24% of its expected value.
A probe's rise time
can be measured with a fast-rising pulse. Since the oscilloscope also has a rise
time
, the measured rise time t needs to be corrected:
All content Copyright © B&K Precision Corporation, except where otherwise noted.
Copying/reprinting/republishing without express written consent prohibited.