Specifications
41
The following picture demonstrates the ability of the scope to average out random noise:
Figure 27
The signal at A was a 200 mV square wave with substantial random noise. At B, the signal has been
averaged 16 times and the noise is substantially reduced. At C, 256 waveforms have been averaged
and the noise is essentially gone.
Averaging can only be used on periodic signals, but since these are often measured, averaging finds
frequent use in day-to-day oscilloscope measurements.
Peak detection
Measuring narrow-width pulses using a slow sweep speed can be challenging with an analog
oscilloscope, even if the scope can trigger on the pulses. A digital scope can use peak detect mode to
show these narrow pulses. Peak detection shows the highest and lowest values from multiple triggers
and thus uses the techniques of equivalent-time sampling. The advantage of peak detection is that it
can show these narrow pulses at slow sweep speeds.
An example of where peak detection is useful is shown in the following picture:
Figure 28
The waveform was a 100 Hz pulse stream with 10 ns wide pulses (this is a 0.0001% duty cycle pulse
train). At A, normal acquisition mode was used and the pulses are not visible, yet the scope had
triggered. At B, peak detection acquisition mode has been turned on and the pulses become visible. An
analog scope would have a difficult time displaying these pulses.
All content Copyright © B&K Precision Corporation, except where otherwise noted.
Copying/reprinting/republishing without express written consent prohibited.