Specifications
35
The AUTO button
One of the biggest time-saving features of a digital oscilloscope is the AUTO button 22. This button tells
the oscilloscope to measure the signals on channels 1 and 2 and display them appropriately. The scope
manual will tell you what signals the scope will be able to measure automatically (or some
experimentation with a function generator will show you). The 254xB scopes require a signal of 50 Hz
or greater, a duty cycle of greater than 0.5%, and an amplitude of at least 10 mVpp to successfully
display a signal using the AUTO button. Knowing your scope's auto measuring capabilities for different
types of waveforms may prove useful in future investigations.
The auto measurement feature is convenient to working engineers and technicians, but may not be
desirable in an educational environment -- the instructor might want the students to learn how to set a
scope up manually. In such cases, the 254xB scopes can have their auto measurement capability
turned off.
Because the AUTO button can be accidentally pressed, a nice feature is to have a menu item to restore
the previous settings. This provides a useful method of measuring a second signal: disconnect signal 1
from channel 1, connect signal 2 to channel 1, press the AUTO button, write down any relevant
measurements from the second signal (or store the waveform), reconnect signal 1 to channel 1, and
press the Undo Autoset menu button to restore the settings you had for signal 1.
Quick help
The 254xB oscilloscopes have a quick help system that provides descriptions of the functionality of
keys and menu selections. Press and hold the key down for one or more help screens describing the
topic. For example, holding down the AUTO button 22 provides the following information on two
screens:
Press this key to get best waveform display by setting vertical, horizontal and trigger controls
automatically. The setups are as follows:
• Display type: Main
• Coupling: DC or AC
• Acquisition mode: Normal
• Vertical scale: Adjust to applicable scale
• Vertical position: Adjust to middle
• Bandwidth limit: Off
• Digital filter: Off
• Invert: Off
• Trigger type: Edge
• Trigger source: Channel with active signal
• Trigger mode: Auto
• Trigger coupling: DC
• Trigger level: 50% amplitude level of the trigger source waveform
• Trig-offset: Reset delay time to zero
• Holdoff: Reset holdoff time to default 100 ns
This quick help feature can be invaluable to a new user to help them understand the various features of
the scope.
All content Copyright © B&K Precision Corporation, except where otherwise noted.