Instruction manual
HB 08-18-2010 8
helpful to refer to previous sections during these exercises. Reminder: the VOLTS/DIV and
TIME/DIV switches are not calibrated unless the knobs in the center of these switches are
fully clockwise in the calibrated positions.
5.1 Preliminaries
Remove any input cables to the scope. Turn on the scope (control 3) and observe LED (1).
In a few seconds you should see a trace which is a horizontal line (it takes the cathode a
few seconds to warm up). If you do not, check your initial control settings. Adjust the
INTENSITY (4) and FOCUS (6) controls for a sharp trace which is easy to view but not
brighter than necessary. Too bright is hard on the phosphor. Move the trace about the
screen with the 2 POSITION controls, 17 (vertical) and 27 (horizontal). Leave the trace
centered. Turn the time base switch (23) to 0.2 s and note the light spot move slowly across
the screen until it gets to the right hand side of the screen. What happens then? Investigate
the effect of turning the variable sweep rate knob (26). Increase the sweep speed one click
at a time and note how the moving spot becomes a solid line. Observe that there is a range
of sweep speeds for which the trace “flickers.” This is due to the persistence of the phosphor
or of the human eye, or perhaps both.
When you need to make measurements of an input signal (peak voltage, period, etc.), it
is useful to use the two position controls (17, 27) to move parts of the trace onto the gridlines
of the screen that have finer divisions.
Try setting the Trigger COUPLING switch (32) to NORMal. The trace should disappear,
as there is no input signal for the scope to trigger on. Set this switch back to AUTO to
retrieve the trace.
5.2 Observing Voltage From The Wavetek FG3C Function Generator
This function generator produces sine, square and triangular waveforms which are selected
by three pushbuttons. The frequency of these waveforms can be adjusted from 0.3 Hz to
3 MHz by means of a knob marked frequency and 7 pushbuttons. The output frequency
appears in the middle of the display panel. At the bottom of the display panel to the right
appear Hz, and in the middle of the bottom there may a multiplier for Hz such as k or
M. It is important to check for the k or M to determine whether the stated frequency is
in kHz or MHz. The multiplier appears and vanishes automatically as you cross certain
threshold frequencies such as 1 kHz. The amplitude of the output is varied by a knob at
the right marked AMPL. The knobs have “in” and “out” positions. The functions for the
knobs are marked in black for the “in” position and blue for the “out” position. In using
this instrument as a simple function generator be sure all the knobs at the bottom of the
front panel are in and that the only two green LED’s that are on are one for the waveform
chosen and one for the frequency range (buttons at top).
Connect the 50 Ω output of this oscillator to the CH1 vertical input (14) of the BK scope
using the coaxial cable. Set the AMPL knob full CCW and turn on both the oscillator and
scope. Adjust the Wavetek for about a 1 kHz sine wave and turn the AMPL knob CW until
you see a reasonable signal.
Vary all of the controls listed below leaving the parameters of the input voltage constant.
Note the effect on the trace of varying a control and understand why the trace changes in
the way it does. In answering the following questions make sketches of a trace if it will help
your explanation. How does the trace change if: