INSTRUCTION MANUAL MANUAL DE INSTRUCCIONES Model 1696/1697/1698 Modelo 1696 / 1697 / 1698 LABORATORY GRADE SWITCHING MODE PROGRAMMABLE DC POWER SUPPLIES with PC INTERFACE
TABLE OF CONTENTS 1. 2. 3. 4. 5. 6. IMPORTANT SAFETY INSTRUCTIONS & PRECAUTIONS FOR USE .............................................................. 3 SPECIFICATIONS ............................................................................................................................................ 4 INTRODUCTION .............................................................................................................................................. 5 CONTROLS & INDICATORS ....................
1. IMPORTANT SAFETY INSTRUCTIONS 1. 2. 3. 4. 5. 6. Do not use this apparatus near water. Clean only with dry cloth. Do not block any ventilation openings. Do not install unit near any heat source or heating emitting devices. Prevent the power cord from being walked on or pinched. Unplug this unit during lightning storms or when unused for long periods of time. PRECAUTIONS 1. 2. 3. 4. 5. The unit must be used within its specified range.
2. Technical Specifications of models 1696, 1697, 1698 Specifications 1696 1697 1698 Output Voltage 1-20VDC 10-5A -40VDC 1-60VDC Output Current 0-10A 0-5A 0-3.3A Ripple & Noise (P-P) 25mV 25mV 25mV < 10 mArms < 10 mArms < 10 mArms Voltage: 0.5% Voltage: 0.5% + Voltage: 0.5% + 200mV 200mV + 200mV Current: 0.2% Current: 0.2% Current: 0.
Weight 6.6lbs. ( 3 Kg) Accessory PC Windows TM software, RS-232 cable and Operation manual 3. INTRODUCTION This series of laboratory grade, switching mode, programmable power supplies are ideal for repetitive test routines in R&D, Production, Product Evaluation and various applications. Information appearing on the large back-lit LCD makes the panel controls simple and easy to use in spite of its sophisticated features.
4. CONTROLS AND INDICATORS FRONT PANEL 1. 2. 3. 4. 5. 6.
4. CONTROLS AND INDICATORS BACK PANEL 7. 8. 9. 10.
5. GENERAL OPERATION PRINCIPLE There are four operation methods for this unit: • Manual Operation • Presets • Timed Program • PC Interface Mode. • The large Liquid Crystal Display will guide you through the various operations/procedures interactively once you understand the operation principle. There are four Dual Function Control Keys for various operation modes and input states of the unit. To change/set operation mode (output on /off, key lock & etc.
Program 1 to 9 is for nine sets of Presets of voltage and current. The program 0 is reserved for the 20 timed subprograms; each is capable of storing 1 second to 99-minute operation period with its preset voltage and current levels. The timed subprograms can be set to operate repeatedly in sequence from 1 to 232 cycles. 6. OPERATING INSTRUCTIONS 6.1 Setting of Operation Modes Output On/Off Mode 1. 2. Press SHIFT key. The display will show Sh:Ft. Press (O/P On/Off) key within 3 seconds.
Upper Voltage Limit Setting This feature limits the upper level setting of output voltage to prevent inadvertent setting of high voltage, which may damage your application. The value of this upper voltage range limit will be retained until further reset. OPERATING INSTRUCTIONS Precaution Reset back to the maximum rated voltage of the unit is recommended to avoid false fault report of the unit 1. Press SHIFT and follow by 0 key. 2. The display will show 3.
OPERATING INSTRUCTIONS Showing output terminals in ON/OFF state. Showing keypad & Jog dial are in Locked (de-activated) or Unlock (activated) state. Unit is ready for or has already been linked up with computer. OUTPUT ENABLE/DISABLE AT POWER UP Press the SHIFT then the UP key. This will enable the output at power up. i.e. When you switch on the power supply, the output is also on automatically with last set voltage value. Press the SHIFT then the DN key. This will disable the output at power up.
STANDARD OPERATION SYMBOLS V- const Constant Voltage mode showing set voltage. V- set This only appears voltage level is ready for setting. I-set This only appears current level is ready for setting. I-const This only appears when unit is in (CC) constant current mode. They show the actual Volts, Amperes and Watts at the output terminals.
6.2 Manual Operation Mode In the Manual Operation Mode, once the voltage and current levels have been set, the unit wi11 keep these settings even after unit has been switched off until further reset. There are 3 ways to input a setting: 1. By the Jog Dial. 2. By the UP or DOWN key. 3. By the Key Pad. For any wrong input, e.g. out of specification values, display will show "OUT OF RANGE" FOR ABOUT 3 SECONDS, then returns to it's previous settings and ready for correct new input.
and returns to its old settings Note : If during any of the above keypad operations, you fall to input a setting within 10 seconds, the unit will cancel the procedure and resume its previous settings. You can exit any of the above operations by pressing the CLEAR key and the unit will revert to its previous settings. 7. USING THE PRESET FEATURES PRESET PROGRAM SYMBOLS Program Unit is ready for inputting of a Preset Program Number (19). Recall Unit is ready for recalling a Preset Program Number.
go to the next Preset. Whenever you want to determine Preset Programming you can press the CLEAR key to return to normal operation mode. Selecting Presets 1. Press the RECALL key The display will show Recall_ 2. Press the number key (1-9). The display will show the chosen Preset Number, its voltage and current level. 3. Press ENTER to activate .The unit will resume the chosen reset values.
9. A subprogram is confirmed to retain the new settings only when procedure 7 is completed. 10. You can always input zero time period as a terminating subprogram. Note: A. B. C. D. Once the subprograms are filled, the unit will return to manual operation mode. Any subprogram with zero time period will become the Terminating Subprogram. For example, if you set the subprogram 7 with zero time period, all the subprograms after 7 (i.e. 8, 9, 10-19) become non-existent.
Note: Beware that the life span of the power supply is limited by the life of the relay, which is l 0 0x 1 0, 00 0 operation as stated by the relay manufacturer. TIMED PROGRAM SYMBOLS 1.
2. Symbols Running in a Timed Program Cycle Unit is ready for the recheck of all subprogram settings.
8. MAINTENANCE 8.1 RECALIBRATION Introduction This in-case recalibration is to reduce the difference between the Set values and the displayed values on the LCD Display. You only use the recalibration when the difference is greater than 0.1V for voltage or -0.01A / +0.02A for current. The whole recalibration for voltages and current takes less than 15 minutes. It is performed by a proprietary software using regression algorithm. The recalibration software is compatible to Window XP, Me, 2000, 98SE, 98.
8.2 TROUBLE SHOOTING 1. Keypad and Dial do not work. Check key lock symbol for lock state, unlock unit by pressing SHIFT then Lock/Unlock key, otherwise switch the unit off and then back on to see if the problem persists. 2. No output power. Check the output for on/off symbol on the display. Press SHIFT O/P ON/OFF key. 3. Cannot get a high voltage setting within the rated maximum. Check the upper voltage limit setting by pressing SHIFT then the 0 key. Reset to the rated maximum voltage. 4.
9. PC INTERFACE CONTROL USER MANUAL 9.1 Introduction We integrate the micro-controller technology and our proprietary software with switching mode power technology to make this unit user friendly, time saving and reliable power supply. In PC interface mode: 1. With our proprietary software one PC can perform remote access/control and output data logging for one programmable power supply with RS-232 or up to 31 programmable power supplies using RS-485 interface. 2.
Installation of software 1. Run Setup.EXE with installation disk that comes with the unit. 2. Follow the instructions in the setup program. During the running of the setup program, you may encounter "VERSION CONFLICT" remarks, ignore it and click yes to complete the installation. 3. Finally a software icon is created in the program menu. 9.3 Operation Instruction for RS-232 Interface 1. Ensure that the PC is off; connect RS-232 to serial COM port of your PC and the power supply. 2.
If it shows No Connection, check the following: A. Go back to setup and check to see if the correct COM port has been assigned. B. Check the power supply; make sure that the RS-232 has been selected. C. Check the RS-232 cable connection. D. Check to make sure that the power supply is turned on. 1. Setting Once you bring up the software you can your desired sample time from 1 second up from the drop menu.
C Min C Max. V Min. W Max. 3. Minimum current level. Maximum current level. Minimum power level in watt. Maximum power level in watt. Log Name Click cursor on "Untitle", type in a name for your log. Note how the ###(5) icon changes to solid color. 4. Log Description You can type in a detailed description of your log. 5. ##### Save Log a) b) c) 6. This function (and the icon) becomes effective when a Log Name is entered to replace the "Untitle".
8. XXX Delete Clicking on this icon will delete the current or retrieve log on display at (3). 9. Print Log in xls format. P.
THE TIME FRAME CONCEPT OF DATA LOG The data logging function starts when the software begins to run. When T Min is set to zero seconds, it means the unit is on real time and the length of time lapsed is on the left hand side of the Time Minimum. T Len is the length of time lapsed strafing from the Time Minimum. Both parameters are adjustable so that any time period of the log can be displayed for analysis.
GENERAL INFORMATION Please refer to (Fig. 3) for the following description. 10. Power Supply Description: serial no S2405000 You may click on and assign an identification number for your power supply in use. Actually this feature is used mainly when you have multiple power supplies in use via the RS-485. 11. Address: This function is for multiple power supply applications. Each power supply has a unique address. Ignore this function when using RS-232. 12. Voltage: ##.
16. OUTPUT Left click on icon will switch on or off the output. Internal Timed Program The PC interface remote mode really eliminates the tedious process in keying in groups of entries on the power supply. Because all the data are displayed together in the monitor, the possibility of wrong entry is greatly reduced. Data of different groups can be classified, store, exported and retrieved for use at any time.
4. 5. 6. 7. 8. 9. Data exceeding the rated voltage and current will not be accepted. Voltage exceeding set UVL (upper output voltage limit) will not be accepted. If retrieved or entered data exceeds upper or lower limit, setting of voltage, current, time the date will become red in color. Transfer set data to power supply by clicking [Save to PS]. Click [Read Fro PS] to initiate the [run] command. Set number of desired [[running cycle] and click [run].
(FIG 4) 3. 4. On your power supplies, press [SHIFT] key, then quickly press [RS-232/485] key and select RS-485. A 3-digit number will be displayed. This number is the address assigned to the power supply and will be used in the software. 5. Using the Key Pad to key in the address to assign for each power supply. The range is 001-031 and each of the power supplies requires a unique address. 6. Switch on your PC and run the SDP program. 7. Click on set up and select the desired Com. Port 1. 8.
9.4.2 Configuration of Multi Windows Analysis 1. 2. 3. In the tool bar, click on Supply Connect, then click on Multi in the drop menu. A multi-window (Fig. 6) will appear. Click on the icon (circled as below) a Multi Power Supply Connect Set Up (Fig.
(FIG. 6) 4. Click on AutoScan Connect, the window will show the connected power supply indicated with “Y”. (FIG. 7) 5. Click on the box along the visible column to set the desired power supply to be visible in Multiple Data log window. 6. You can type in the description and location of your power supplies in the location and description column. 7. Click on Close Icon (at right bottom of window) to return to Multiple Data Log Window.
8. Remarks: 33 1. Multi Alleyway Display one click and it will display the Data Log and output data of all the power supplies. It will activate the icon 2, 3 & 4. 2. Show Digital and Log one click and it will show both the data log of all connected power supplies.
(FIG. 9) You can click on the data log to select the power supply, the data log will highlight in blue and the address bar in the left bottom window will show the selected power supply. 3. Show Digital - one click and will show the digital readings of all the connected power supplies. 4. Show Log - one click and it wi11 show the data log of all the connected power supplies. 5. Single Alleyway Displays - one click and it will only display the Data Log of the selected power supply (Fig. 10).
6. Log Thumbnails Size Set Up - one click allows user to adjust the window size of the Data Log Window in Multi Alleyway Display. Use the sliders to adjust the height and width of the Data Log Windows. Save 4:3 tick box can enable 4:3 screen size for the Data Log Windows. 7. Save Log: a. b. c. d. e. f. Click on it and save log window (Fig. 13) will appear. Click on the box along the Save column to choose the desired power supply data log to save. Type in the table name.
10. Delete Log It will delete log data in PC 11. Export to a file log of xls type Click on this icon to export the collected data in Data Log in xls format to your PC. 12. Open File Log of xls type Click on this icon to export the collected data to the SDP software. (FIG. 12) 13. Close File Log of xls type Click this icon to close the import xls format file. 14. Print Log Prints log into xls format. 15. Log Click on 16. Sample Click to select the sampling time 17.
Service Information Warranty Service: Please return the product in the original packaging with proof of purchase to the address below. Clearly state in writing the performance problem and return any leads, probes, connectors and accessories that you are using with the device. Non-Warranty Service: Return the product in the original packaging to the address below. Clearly state in writing the performance problem and return any leads, probes, connectors and accessories that you are using with the device.
Limited One-Year Warranty B&K Precision Corp. warrants to the original purchaser that its products and the component parts thereof will be free from defects in workmanship and materials for a period of one year from date of purchase. B&K Precision Corp. will, without charge, repair or replace, at its option, defective product or component parts. Returned product must be accompanied by proof of the purchase date in the form of a sales receipt. To obtain warranty coverage in the U.S.A.
22820 Savi Ranch Parkway Yorba Linda, CA 92887 www.bkprecision.com © 2006 B&K Precision Corp.