Manual
30 
1.  The following instructions are for use by qualified 
service personnel only. To avoid electrical shock, do 
not perform servicing other than contained in the 
operating instructions unless you are qualified to do so. 
2.   When the unit is plugged into an ac outlet, even if the 
POWER ON switch is off, ac line voltage is present as 
some input power circuits.  Observe caution any time 
the case is removed from the instrument. 
PREVENTIVE MAINTENANCE 
  Do not overload the instrument. Excessive output current can 
damage the variable auto-transformer. Prolonged use at high 
current accelerates failure. Keep output current as low as 
possible, and never exceed the maximum rating of the unit. If 
possible, power only one piece of equipment at a time. Use the 
ISOLATED OUTPUT only for the equipment under test-not for 
other test equipment. It is very important to observe the 
maximum current derating (Table 1) above 130 volts, and to 
allow sufficient cooling time when operating Model 1655A at 
greater than 3 amps outlet. 
  Periodic inspection of the brush on the variable contact of the 
auto-transformer is recommended.  Badly worn or arcing brushes 
should be replaced. See the AUTO-TRANSFORMER BRUSH 
REPLACEMENT paragraph for replacement procedures. 
RETURNING FOR SERVICE 
  If you have encountered a problem with your unit and choose 
to return it for service, check the following before returning. 
1.  Is the circuit breaker reset (Model 1655A)? See FUSE 
REPLACEMENT paragraph below if required. 
2.  Are the fuses okay? See FUSE REPLACEMENT 
paragraph below if required. 
FUSE REPLACEMENT 
  The ac power supply will trip its circuit breaker or blow a fuse 
only if it is severely overloaded or a failure occurs in the 
instrument. Always turn off the unit to reset the circuit breaker or 
replace a fuse. If fuses are internal, unplug the instrument from 
its wall outlet before attempting replacement. Before turning the 
unit back on, reduce the output voltage or remove the overload 
that caused the blown fuse. Replace the fuse only with the 
original type and value-never one of a large value. A larger value 
may create a fire and safety hazard, or cause serious damage to 
the equipment. Also refer to the following specific fuse 
replacement information for each model. 
Model 1653A 
  If the instrument is severely overloaded at higher output 
voltages, fuse F1 on the rear panel will blow. The unit will stop 
operating and the pilot light will go off. If the instrument is 
severely overloaded at low output voltage, internal fuse F2 will 
blow. In this case, there will be no output but the pilot light will 
stay on. Fig. 14 shows the location of the fuses. Replace fuse F1 
only with a 3-amp slow-blow type 3AG. 
MAINTENANCE AND CALIBRATION
WARNING










