Reset Key enabled, Channel B Pulse Counter
Reset key enabled Reset key disabled
If you have not enabled the reset key, the stringTOTAL
is displayed.
Having completed the programming, disconnect the
programming input PROG from 0 V.
Please note that only the parameters confirmed with S/E
are accepted.
6. Technical Data
Displays
Special LC display with dimension line, 8 digits, 10 mm
digit height, pre-decimal point zero suppressed.
Display Capacity: Rate meter 99999999
Automatic decimal point
Pulse counter 99999999
Programmable decimal point
Accuracy
Period measurement accuracy:
Programmable to 0.1%, 1%, or 10% (corresponds to a
minimum resolution of 4, 3, or 2 digits, respectively).
The world of flow
Seite 18
Power Supply
internal lithium battery 3.6 V / 1.2 Ah
Average battery life 5 years
LED Display backlighting
The LED display backlighting must be operated with
external voltage connected to 24 VDC and 0 V screw
terminals.
External Voltage: 24 VDC max. residual ripple 5%
absolute limits 19 to 30 VDC
Electromagnetic Compatibility (EMC)
Interference emission EN 55011 Group 1 Class B
Interference strength EN 50082-2
EN 61010-1 Measuring Insulation Voltage
100 Veff, Contamination Class 2, Surge Category III
DIN VDE 0411 Protection Class
Protection Class II
Electrical Connection
Terminal screw connection, P Phillips screws, size 1
max. lead section 2 x 1.5 mm²
min. lead section 2 x 0.2 mm²
IEC 529 Protection Class
IP 65 front
Temperature / Humidity range
Operating temperature range - 10°C to + 50°C
Storage temperature range - 20°C to + 70°C
Temperature / Humidity 90% relative humidity @ 38°C