User manual

OPERATION MANUAL
Technical changes reserved
0141 0316-196 27.08.2015
Humidity- temperature probe
with USB-Interface
WINDOWS-Software RECORDER
With the help of this program, which is covered in the scope of supply,
the measured values can be received through the USB-interface and
displayed on the PC. The displayed le is compatible with any desired
spreadsheet program, with which it is possible to further process,
statistically evaluate or visualise the measurement data.
In addition, the PC-Software also calculates dew point, absolute
humidity, enthalpy and vapour pressure from the measured values of
relative humidity and temperature. The calculated gures can also be
stored.
View of the Software “RECORDER“
System requirements
Windows 98, 2000 or XP, RS232 or USB-interface. Generally, older
PCs are also suitable.
Important hint: First connect the USB-Version to the PC after
installing the software. This simplies driver installation and enables
“Plug&Play“ feature.
Installation
A detailed installation instruction is provided on the CD, which
automatically gets started on inserting the CD (prerequisites: Internet-
Explorer 5.0 or higher). Follow these instructions for installation:
Manual Installation
Insert the enclosed CD into your drive and select „Run“ in the start-
menu and then browse to select the le ‚setup.exe‘ under the path
LW:\software\RECORDER\TEMPLOG\disk1. Then follow the instruc-
tions of the installation program.
First time operation
Connect the humidity meas-uring system to the USB-interface of a
PC. After rst time run of the software, go to menu option „settings“
and select device type as „HYTELOG 4800Bd“ and also select the
type of interface to be used under „connections“ (Note: For USB-ver-
sion, mention the virtual COM-port specied during driver installation).
The remaining settings (Data rate, Parity, Start and Stop bit) are
automatically selected and need not be changed. If the connection is
established, the data communication appears on the terminal window.
Then select „Close“. The current settings will be stored.
If you are not able to establish data link between PC and the
measuring device or the measuring adaptor, then rst please check
the power supply and also cable connection to the PC. Further in-
formation on debugging is available under FAQ´s on the CD or at our
Homepage under SUPPORT.
Data recording
First activate all the hooked up measurement channels that are to be
recorded. In ‚Text 1‘ and ‚Text 2‘, you can enter a description, which
has to appear as heading on the top of data le. The data is recorded
in a le, which is de-clared as path in the ‚Start‘ button. The recording
begins with the ‚Start‘ button.
EXCEL™
The created le is compatible with CSV-Format. In order to display
the measured data, you can use graphic tools, for example, the
diagram-assistant. However, other programs can also be used to
graphically represent or evaluate the measured data.
Calibration
The humidity probe is supplied in calibrated
condition. The accuracy at 23°C is of the order
of ±0.3° K and ±2 % RH. Under normal opera
ting conditions, it is not necessary to again
calibrate the probe. The cross checking of
measuring accuracy of the humidity measu
rement part can be done by end user with the
salt reference cells available as special ac
cessories.
The cross checking must be done in temperature stable environment.
In case, the measured value does not match with the reference value
of the cell, the device should be sent to our calibration laboratory for
check-up.
Humidity- temperature probe
with USB-Interface
B+B Thermo-Technik GmbH | Heinrich-Hertz-Straße 4 | D-78166 Donaueschingen
Fon +49 771 83160 | Fax +49 771 831650 | info@bb-sensors.com | bb-sensors.com
2 / 4