Service manual

© Adam Equipment Company 2007
5
Off Centre Loading error Overload Stops not correct
Obstruction under pan
Battery will not charge Charging circuit failure
Battery Failure
Incorrect adapter (below 800 mA)
Processor failure
4.0 ERROR CODES
ERROR
CODE
DESCRIPTION POSSIBLE CAUSES
Err 1*
Time input Error Invalid time entry such as “268970” for
the time format “
H-m-S”.
Err 2*
Date input Error 34
th
day of a month is an invalid entry.
Err 4
Initial Zero is greater than
allowed (4% of maximum
capacity) when power is turned
on or when the [Zero/Enter]
key is pressed.
Weight on the pan when turning the
scale on.
Excessive weight on the pan when
zeroing the scale.
Platform is not installed.
Improper calibration of the scale.
Damaged load cell.
Damaged Electronics.
Err 6
A/D count is not correct when
turning the scale on.
Load cell is damaged.
Electronics is damaged.
Err 7
Percent input error Percent function is entered with no
reference mass on the pan.
Err 8
High limit input error Low limit is set first, then the high limit is
set lower than the low limit and high
limit not equal to zero.
Err 9
Low limit input error High limit is set first, then the low limit is
set higher than the high limit and low
limit not equal to zero.
FAIL H or
FAIL L
Calibration error Improper factory calibration). The old
calibration data will be retained until the
calibration process is complete.
*Err 1 and Err 2 are not applicable for scales used in approved applications.