Service manual

© Adam Equipment Company 2007
15
8.7 F7 -A/D COUNT
To enter this parameter, press the [Tare] key when “
F7 Cnt” is shown.
This parameter allows you to view the A/D counts from the internal A/D converter. This can
be an aid to service.
Press the [Tare] key to return to the PARAMETER menu.
Typical value at zero is 30,000-90,000 (approx.)
Typical value at full capacity is 500,000 (approx.)
8.8 F8 –ZERO
This parameter allows the normal method of setting zero.
To enter this parameter, press the [Tare] key when “
F8 ZE M” is shown.
The display will show the current settings, “
modE 1” or “modE 2”.
Press the [Func] key to scroll through the options.
Press [Tare] to accept the displayed value and return to the parameter or press the [Zero]
key to return to the parameter without changing the value.
Then press [Func] to advance to the next parameter.