Suggested Calibration Weights:
Calibration weight 1 Calibration Weight 2
CBK 3M 1 kg 2 kg
CBK 6M 2 kg 4 kg
CBK 15M 5 kg 10 kg
CBK 30M 10 kg 20 kg
8.2 F2 -RESOLUTION
NOTE: This value can not be changed for the CBK-M scales.
8.3 F3 – CAPACITY
To enter this parameter, press the [Tare] key when “F3 CAP” is shown.
The display will show the current capacity.
Press the [Func] key to scroll through the options.
Press [Tare] to accept the displayed value and return to the parameter or press the [Zero]
key to return to the parameter without changing the value.
Then press [Func] to advance to the next parameter.
NOTE: The scale must be originally built for the capacity selected. The 3 kg unit uses the
5 kg load cells, 6 kg unit uses 10 kg load cells, the 15 kg unit uses 20 kg load cells and the
30 kg unit uses 35 kg load cells.
8.4 F4 -INITIAL ZERO RANGE
To enter this parameter, press the [Tare] key when “F4 int” is shown.
The display will show the current initial zero range.
Press the [Func] key to scroll through the options.