Operating instructions

Commissioning
FLOWSIC600 · Operating Instructions · 8010125 V 4.0 · © SICK AG 65
Subject to change without notice
4.4 Identification
4.4.1 Checking identification, operation / design data and firmware version
Before commissioning, cross-check the data representing the flow meter with the data in
the test protocols which are contained in the Manufacturer Data Report (MDR). This can be
done on the LCD display (see Technical Information) or - much easier - with
MEPAFLOW600 CBM software:
Open the "Meter information" page and compare the data in the "Identification" section
(
Figure 23) with the data in the check reports of the MDR or, when the meter has been
calibrated, with the calibration and parameter report.
Figure 23 "Meter Information" page
"Identification" section