Operating instructions
Aztec ADS430 | Optical dissolved oxygen probe | OI/ADS430–EN Rev. A 7
4 Sensor setup – first-time installation
To perform a first-time installation (Easy Setup menu):
1. Connect a new or replacement probe to the transmitter’s
EZLink connector – see transmitter Operating instructions
OI/AWT440-EN.
The New Sensor S(1 – 4) Detected Start Easy Setup ?
prompt is displayed identifying the new / replacement
sensor (S1 to S4):
2. To enter Easy Setup level, press the key
(below the icon).
The Easy Setup start screen is displayed:
Press the key (below the Select prompt).
3. To enter Easy Setup level, press the key
(below the Select icon).
The Configuration parameter is displayed on the left of the
screen and the factory default value for it is displayed on
the right.
4. Press the key (below the Edit prompt) to change the
default value to the required value / selection.
5. Press the key (below the Next prompt) to accept the
value / selection displayed and advance to the next
configuration parameter.
The following Configuration parameters are set at Easy
Setup level:
—Tag
—PV Type
—Units
— Range High
—Range Low
—Clean Interval
— Salinity Units
— Salinity Correction
— Pressure Units
— Barometric Pressure
6. Continue with configuration of the required parameters.
On completion the Easy Setup start screen is displayed:
7. To exit Easy Setup, press the key (below the Exit
prompt) to display the Operator Page.
Pressing the key (below the Select prompt) re-enters
the Easy Setup level where parameters can be reviewed or
modified after 1st time connection.
After completing the Easy Setup level, pressing the
or key enters the Advanced Configuration level, where
all available sensor and transmitter parameters can be
reviewed or modified.
8. When the probe has been configured, perform a Calibrate
routine – see Section 6, page 11.
IMPORTANT (NOTE)
Perform this procedure when a new / replacement
sensor cap is connected to the transmitter for the first
time only. For existing sensors, see Section 5, page 8.
AWT440
New Sensor S3
Detected
Start Easy
Setup ?
00-00-2014
15:21:24
Menu
Easy Setup
Exit Select
IMPORTANT (NOTE)
Refer to Section 5.1, page 9, for parameter details – not
all parameters in Section 5.1 are displayed at Easy
Setup level.
IMPORTANT (NOTE)
To re-configure an existing probe (after first-time
installation), enter the Configuration level (see Section
5.1, page 9) via the Operator Page – refer to transmitter
Operating instructions OI/AWT440-EN for Operator
Page details and navigation.
Menu
Easy Setup
Exit Select