Operating instructions
Aztec ADS430 | Optical dissolved oxygen probe | OI/ADS430–EN Rev. A 11
6 Calibration
This section describes how to calibrate the probe and
involves measuring the probe's sensitivity to oxygen by
exposing the probe to water-saturated air.
Calibrations are initiated via the Cal prompt displayed on
Operator pages or via the Calibrate and Advanced menu
items on the Access Level page – refer to transmitter Operating
instructions OI/AWT440-EN for all transmitter menu options.
6.1 Calibrate menu
IMPORTANT (NOTE)
— Do not perform a calibration until the probe and
transmitter are installed and ready for operation.
— Before removing the probe for calibration purposes,
set the currents outputs and alarms to Hold
(enabled via the Operator Menu / Manual Hold
function).
— After calibration and probe replacement, reset the
currents outputs and alarms – see Section 6.1
below.
Used to calibrate the probe.
Access to the Calibrate menu is permitted via the Calibrate and Advanced levels only.
Note. During calibration, current outputs and alarms are set to Hold automatically if
Hold Outputs is enabled (see below).
Menu
Calibrate
Exit
Select
Menu Comment Default
S1(to 4) :RDO Select the optical dissolved oxygen probe to calibrate.
Probe Calibration
Calibration Type
1-Point Cal Select to perform a 1-point calibration in water-saturated air
(100% saturation).
2-Point Cal Select to perform a 100% saturation calibration in water-saturated
air and a 0% saturation calibration in an oxygen-depleted solution,
e.g. sodium sulphite (Na
2SO3).
Hold Outputs Enable / disable the Hold Outputs function. If enabled, the current
outputs and alarm functions are held during calibrations.
Enabled