User Manual
AXONICS SNM SYSTEM® – Neurosmulator | Rechargeable, Model 5101
7
SPECIFICATIONS
Table 1 shows the Neurosmulator physical specicaons. For detailed descripons and specicaons for other
components and accessories, refer to the product literature for those devices.
Table 1: Neurosmulator Specicaons
Physical Aributes Height 42 mm
Length 22 mm
Thickness 6 mm
Weight 11 g
Volume 5 cc
Radiopaque idener GRJ
Smulaon Characteriscs Frequency 2 – 50 Hz
Pulse Width 60 – 330 µs
Amplitude 0 – 8 mA
Minimum Amplitude Step Size 0.05 mA
Ramping 0 – 30 s
Smulaon Mode Connuous or Cycling
Mode of Operaon Current-Controlled
Power Source Baery Rechargeable
Note:
All dimensions are approximate.
Table 2 shows the Neurosmulator baery recharge interval at various energy sengs.
Table 2: Baery Recharge Interval
Amplitude (mA) Rate (Hz) Pulse width (μs) Recharge Interval
0.1 mA 14 210 10 months
1 mA 14 210 8 months
1.5 mA 14 210 6 months
2 mA 14 210 5 months
Note: Sengs include connuous smulaon at 1200 Ohms impedance
Table 3 shows the materials used in the Neurosmulator kit components that come in contact with human ssue.
Table 3: Human-Contact Materials
Device Component Material
Neurosmulator Neurosmulator case Titanium-Ceramic
Neurosmulator header Epoxy
Septum and strain relief Silicone
Setscrew Titanium
Adhesive Silicone
Torque wrench Torque wrench handle Polyetherimide
Torque wrench sha Stainless steel
Note:
The Neurosmulator case, which contains the electronics and power source, is hermecally sealed.