User Manual

AXONICS SNM SYSTEM® – Neurosmulator | Rechargeable, Model 5101
6
STORAGE AND USAGE ENVIRONMENT
Component Packaging
Any component that has been compromised in any way should not be implanted; this includes the re-use of non-reusable
product. Do not implant the component if any of the following have occurred:
The storage package or sterile pack has been damaged, pierced, or altered, as sterility cannot be guaranteed, which may
lead to infecon.
The component itself shows any signs of damage. The component may not funcon properly.
The use-by date has expired. In this case, component sterility cannot be guaranteed, and infecon may occur.
The sterile component was dropped onto a non-sterile surface. In this case, the sterility cannot be guaranteed, and
infecon may occur.
Usage Environment:
The following lists the appropriate temperature, humidity, and pressure usage condions for use of the Neurosmulator:
Temperature: 20 °C to 45 °C
Pressure: The Neurosmulator should funcon at up to 10 m (33 feet) underwater (200 kPa) and at altudes up to
3000 m (10,000 feet) associated with acvies like hiking and skydiving (as low as 70 kPa)
Shipping and Storage Environment:
The following lists the appropriate temperature, humidity, and pressure condions for shipping and storing the Neurosmulator:
Temperature (short term: 3 days): -10
o
C to 55
o
C
Humidity (short term: 3 days): 15% to 95%
Pressure (short term): 57 kPa to 106 kPa
If the
Neurosmulator is exposed to extreme temperatures, it may be permanently damaged and should not be used, even
if it has returned to a temperature that is within the specied operang range.
Sterilization
The contents of this package have been sterilized using ethylene oxide. This device is for single use only and should not be
re-sterilized.