User Manual
AXONICS SNM SYSTEM® – Neurosmulator | Rechargeable, Model 5101
12
2. Use the Clinician Programmer to check the impedances and ensure good funcon and connecvity of the system.
Notes:
• The Neurosmulator should be in the subcutaneous pocket during system interrogaon to ensure proper readings.
• Refer to the Clinician Programming Manual for detailed instrucon on checking the system integrity and impedances.
3. Use the suture hole in the header to secure the Neurosmulator to the muscle fascia with non-absorbable silk.
Completing the Implant Procedure
1. Close and dress all incisions.
2. Any opened or used products which are no longer sterile should be discarded appropriately. It is recommended to
recycle non-biohazardous waste to the extent possible.
3. Program the paent’s Neurosmulator and Remote Control. Refer to the Clinician Programming Manual for more
detailed instrucon.
4. Give a Remote Control and paent ID card to the paent.
Cauon
The paent should carry the Remote Control at all mes to be able to adjust or turn o the Neurosmulator.
5. Complete the system registraon paperwork and return to Axonics.
6. Schedule the paent’s follow-up visits at regular intervals to ensure that the smulaon is programmed opmally.
Post-Surgery Management
Paents should be monitored post implant. Administer prophylacc anbiocs for 24 hours.
Paents should avoid sudden movements and pung pressure on implant site. This will reduce risk of movement or
migraon of the implanted system.
Replacing the Neurostimulator
1. Carefully open the implant site and remove the Neurosmulator from the subcutaneous pocket. Avoid cung the ned
lead to preserve for connecon with the new Neurosmulator.
2. Clean the Neurosmulator connector block and lead with sterile water. Wipe both dry with sterile gauze.
3. Use the torque wrench to loosen the setscrew in the Neurosmulator connector block by turning it counterclockwise
(Figure 5).
4. Gently remove the lead from the Neurosmulator.
Cauon
Replace any device that shows signs of damage, ping, or corrosion.
5. Set aside the explanted components.
6. Connect the lead and replacement Neurosmulator according to the steps above.
Dispose of explanted devices as bio-hazardous waste.