User Manual
AXONICS SNM SYSTEM® – Neurosmulator | Rechargeable, Model 5101
11
• Avoid pulling the lead body taut when implanted.
• Do not aempt to insert the lead into the Neurosmulator if the setscrew is not suciently
retracted as doing so may cause damage to the lead and/or cause the lead to not seat fully into
the connector block.
• Ensure that the setscrew ghtens on the retenon sleeve, not an electrode. Tightening the
setscrew onto the contact could damage the contact, leading to lack of therapy.
5. Fully insert the torque wrench into the hole of the Neurosmulator connector block. Tighten the setscrew by turning
the torque wrench clockwise unl it clicks (Figure 6).
Figure 6: Secure the Lead by Tightening the Setscrew Clockwise onto the Retenon Sleeve.
• Ensure that the torque wrench is fully inserted into the setscrew. Otherwise, the setscrew may be
damaged, which can result in intermient or loss of smulaon.
• The torque wrench is designed for single use only and cannot be assured to work appropriately if
used for mulple surgeries. Discard the torque wrench aer use.
Implanting the Neurostimulator
1. Place the Neurosmulator into the subcutaneous pocket. Ensure that the ceramic side is placed away from the paent’s
midline to ensure good communicaon with the Remote Control and ease of recharging (Figure 3). The etched wring
can face either towards or away from the muscle ssue. Ensure that the lead curves gently away from the Neurosmu-
lator with no sharp bends.
Note: The Neurosmulator should be placed no deeper than 3.0 cm (about 1 in) below the skin and should be parallel
to the skin. If the Neurosmulator is too deep or is not parallel to the skin, telemetry and/or charging may be unsuccessful.
Cauons
Do not coil excess length in front of Neurosmulator. Wrap excess length around the perimeter of the
Neurosmulator (Figure 7) or place under the Neurosmulator to minimize interference with telemetry
during programming.
Figure 7: Wrap Excess Lead Around or Under, but not on Top of, the Neurosmulator.
Cauons
Cauons