User Manual
AXONICS SNM SYSTEM® – Neurosmulator | Rechargeable, Model 5101
10
• The Neurosmulator implant site should be irrigated with anbioc soluon, and it is
recommended that IV anbiocs be administered perioperavely. Do not soak the
Neurosmulator in anbioc soluon as this may aect lead connecons.
• The Neurosmulator has been sterilized. The Neurosmulator should not be placed on
any non-sterile surface. The Neurosmulator should not be placed on skin. An infecon
may require surgical removal of the implanted system.
2. Use the tunneling tool to create a tunnel from the lead incision site to the neurosmulator pocket. Refer to the Tined
Lead Manual for detailed tunneling and lead implant instrucons.
Connecting the Lead to the Neurostimulator
1. The components should be wiped and dried to remove any uids before making the connecons. If necessary,
use sterile water or a non-ionic anbioc soluon, then wipe dry.
Cauon
Failure to completely dry the components could lead to undesired smulaon, intermient
smulaon, or loss of therapy.
2. Ensure that the Neurosmulator connector block is dry and clean.
3. Use the torque wrench to turn the setscrew counterclockwise to back up the setscrew. Do not remove the
setscrew from the connector block (Figure 4).
Figure 4: Use the Torque Wrench to Turn the Setscrew Counterclockwise to Back up the Neurosmulator Setscrew and
Allow for Inseron of the Lead.
4. Insert the lead into the Neurosmulator connector block unl fully seated and the lead cannot be inserted further.
Marker D on the lead should be inside the Neurosmulator strain relief (Figure 5). The retenon sleeve on the ned
lead should be posioned under the Neurosmulator setscrew.
Figure 5: Insert Lead Fully into the Neurosmulator Connector Block.
Cauons