User manual
85
4.5.1 Signal Averaging
If Signal Averaging is used, each sweep is first acquired and plotted in Gray, then the average of the
sweeps is then plotted in LightBlue. The setting for the number of sweeps to average for the Slow Repeat
Sweeps is set by the AvgLoop number of loops field - see Fig. 3.1.2.1 for a simple averaging protocol,
and Figs. 7.4.1.1, 7.4.9.1 and 7.4.10.1 for more complex averaging protocols.
You can also signal average during Fast Repetitive Pulse Sweep (LTD) Stimulation which is set by the
NumSweeps to Avg field (Fig. 4.4.2.1).
4.5.2 Stimulus Artifact Blanking
Fig 4.5.2.1 and Fig 4.13.2.1 show the effect of stimulus artifact blanking. If Stimulus Artifact Blanking is
chosen, stimulus artifact blanking is enabled, and a S0 or S1 Blank field appears on the Detection page
(see Fig. 4.5.2.1). Beginning at the time of the start of the S0 or S1 pulse to the time set in the S0 or S1
Blank field (1 ms in Fig. 4.5.2.1), the stimulus artifact data in the ADsweep array is (sometimes) set to the
Average of the 1 data point value just before blanking begins, and one point value just after blanking
ceases (Fig. 4.5.2.1a). If the time in the S0 or S1 Blank field is set too short, then the stimulus artifact is
not entirely blanked, if it is set too long (but not ridiculously long) generally there aren't any problems as
long as the peak amplitude is not clipped.
WinLTP can take the Average of the point just before blanking begins and the point just after blanking
ceases (Fig. 4.5.2.1a). However, WinLTP can also take the Slope between the point just before blanking
begins and the point just after blanking ceases (Fig. 4.5.2.1b), and can also take the point just before
blanking begins and then Hold that voltage until to the point just after blanking ceases (Fig.
4.5.2.1c). Setting the Average, Slope or Hold blanking method is accomplished by entering A, S or H
into the Average/Slope/Hold field next to Blank.
Stimulus artifact blanking is useful for determining the peak amplitudes of individual EPSPs when the
stimulus artifacts are riding on top of the previous EPSP, or when trying to determine the area or peak
amplitude of a whole train which could be seriously distorted by the stimulus artifact (Fig 4.13.2.1).
Stimulus artifact blanking has also turned out to be useful when trying to fit exponential curves to the
decay phase of closely spaced EPSCs when the artifact for the next EPSC occurs during this decay
phase. This approach has been used for exponential curve fitting of synaptic trains by the Synaptosoft
MiniAnalysis Program (www.synaptosoft.com). It's interesting how removal of stimulus artifacts allows
EPSPs and EPSCs to be analyzed as spontaneous events. The Slope method has proved particularly
useful for fitting exponential curves to synaptic potentials by having a smoother decay phase; in fact it is
almost impossible to see where the stimulus artifact was.
The Hold method has proved useful for blanking stimulus artifacts where the point just after blanking
ceases varies widely (such as blanking the stimulation in extracellular CA3 recording where the ntidromic
spike occurs right after the stimulus artifact).