APPENDIX B Limitations to WinLTP
B.1 You can Reanalyze no more than about 2000 files at one time
This is due to a bug in the AmpFile -> Reanalyze… “Select Files to reanalyze” dialog box, which can
select only about 2000 files at one time.
To work around this bug open the AmpFile -> AutoClear Analysis Graphs for Next Reanalysis…
“AutoClear for Next Reanalysis” dialog box and uncheck the CheckBox:
[ ] AutoClear the Analysis Graphs and Spreadsheet for Next Reanalysis
Then do repetitive reanalyses of up to about 2000 files each, e.g. five repeats for 10,000 files reanalyzed.
Hopefully, this limitation will be removed in the not to distant future.
B.2 Maximum duration MainProtocol can be run with M-series
boards is 14hrs, 54min
This limitation can now be overcome by using Protocol Linking (Chapter 9).