User manual
226
If “0:30 AP5” was manually entered by keyboard input AFTER the AddedEvent time of 0:30 at about 42
seconds after the MainProtocol started, the result in Fig 11.7.1c occurs.
16:39:46.0 0.0 Start MainProtocol
16:39:46.5 0.5 Enter Loop0
16:39:46.5 0.5 0N250070.P0, First Loop sweep
16:39:51.5 5.5 0N250071.P0
16:39:56.5 10.5 0N250072.P0
16:40:01.5 15.5 0N250073.P0
16:40:06.5 20.5 0N250074.P0
16:40:11.5 25.5 0N250075.P0
16:40:16.0 30.0 AP5
16:40:16.5 30.5 0N250076.P0
16:40:21.5 35.5 0N250077.P0
16:40:26.5 40.5 0N250078.P0
16:40:31.5 45.5 0N250079.P0
16:40:36.5 50.5 0N250080.P0
16:40:41.5 55.5 0N250081.P0
16:40:46.5 1:00.5 0N250082.P0
16:40:48.9 1:02.9 Stop MainProtocol
Fig. 11.7.1.c. Experimental Log output if “0:30 AP5” was entered AFTER the AddedEvent time of 0:30
seconds, (e.g. at 0:42 seconds after start of MainProtocol) , and P0sweep was checked.
However, in WinLTP 1.11 and subsequent versions, if you manually enter a solution change by keyboard
input BEFORE or AFTER the first sweep following the solution change time, then this first sweep following
the solution change time is automatically printed to the Experimental Log irregardless of whether that
sweep checkbox was checked (Figs. 11.7.1.d and e). This is true for entering a solution change by
keyboard input AFTER the first sweep following the solution change time as long as fewer than 5000
sweeps have occurred in between (because WinLTP remembers the time of only the last 5000 sweeps).
16:47:27.0 0.0 Start MainProtocol
16:47:27.5 0.5 Enter Loop0
16:47:27.5 0.5 0N250099.P0, First Loop sweep
16:47:57.0 30.0 AP5
16:47:57.5 30.5 0N250105.P0, First sweep after AddedEvent
16:48:30.6 1:03.6 Stop MainProtocol
Fig. 11.7.1.d. Experimental Log output if “0:30 AP5” was entered BEFORE the AddedEvent time of 0:30
seconds, (e.g. at 0:12 seconds after start of MainProtocol), and P0sweep was NOT checked
16:51:44.0 0.0 Start MainProtocol
16:51:44.5 0.5 Enter Loop0
16:51:44.5 0.5 0N250115.P0, First Loop sweep
16:52:14.0 30.0 AP5
16:52:14.5 30.5 0N250121.P0, First sweep after AddedEvent
16:52:47.5 1:03.5 Stop MainProtocol
Fig. 11.7.1.e. Experimental Log output if “0:30 AP5” was entered AFTER the AddedEvent time of 0:30
seconds, (e.g. at 0:42 seconds after start of MainProtocol), and P0sweep was NOT checked